6.2.2.1.1 - standard operation (IoC Disabled and Gate step Disabled)
Description
The external trigger generates one gate pulse per exposure. A gate delay is applied that is the same for each exposure
in the series.
Setup
IOC disabled, gate step disabled, output A/B/C disabled.
Waveforms
Operation
External Trigger
Fire
gate Monitor
ARM