AD-4430B
- 49 -
7.6.4.
The Hold Function
(
)
Function No.
Parameter
and range
Item name
Description
Default
value
User
setting MEMO
0.00 to 9.99
Average
time
Time to perform the averaging. Unit
:
second.
0.00 is not averaged.
0.00
0.00 to 9.99
Start wait
time
Time to wait before commencing hold or averaging.
Unit
:
second.
0.00
0 to 2
Condition of
automatic
start
Condition to start the hold or averaging automatically.
0
:
Do not use automatic start.
1
:
Above the near-zero range, and stable.
2
:
Above the near-zero range.
0
0 to 1
Release
using
control input
Release when control input is falling.
0
:
Do not release
1
:
Release
1
0.00 to 9.99
Release
time
Release after a set amount of time has passed.
Unit : second. 0.00 is not released.
0.00
0 to 99999
Release
using
fluctuation
range
Release when fluctuation from the holding value
exceeds a set value. The decimal point position is
the same as the setting of
.
0 is not released.
0
0 to 1
Release at
near-zero
Release when the weighing value is in the near-
zero range.
0
:
Do not release.
1
:
Release.
0
7.6.5.
The BCD Function
(
)
Function No.
Parameter
and range
Item name
Description
Default
value
User
setting MEMO
1 to 4
Data
outputting
1
:
Weighing value
2
:
Gross weight
3
:
Net weight
4
:
Data specified at BCD input
1
1 to 3
Data
transfer
mode
1
:
Stream
2
:
Automatic printing
3
:
Manual printing
1
5 to 1000
Data
transfer
rate
5 times/second
10 times/second Display linked to
20 times/second
100 times/second
1000 times/second
20
1 to 2
Output logic
of data
transfer
1
:
Negative logic
2
:
Positive logic
2
1 to 2
Output
logic of
negative
sign
1
:
Negative logic
2
:
Positive logic
2