![AND AD-4430B Скачать руководство пользователя страница 30](http://html1.mh-extra.com/html/and/ad-4430b/ad-4430b_instruction-manual_2941312030.webp)
- 28 -
AD-4430B
Function No.
Parameter
and range
Item name
Description
Default
value
0 to 1
Zero setting when
power is turned on
Select whether or not to perform zero setting when
power is turned on.
0
:
Disables.
1
:
Enables.
0
-7.0000
to 7.0000
Input voltage at zero
Input voltage from a load cell at zero. Unit
:
mV/V.
This value is determined in zero calibration during
the calibration with an actual load.
0.0000
0.0001
to 9.9999
Input voltage at span
Input voltage from a load cell at span. Unit
:
mV/V.
This value and the value of
are determined
in span calibration during the calibration with an
actual load.
3.2000
1 to 99999
Calibration weight
corresponding to
Input voltage at span
The input voltage at span for
displays the
input voltage per this setting of weighing value.
When performing digital span,
,
and
are required. The decimal point position is
the same as the setting of
.
32000
NOTE:
*1
Record the setting values of
,
and
in the “Setting List” at the end of
the manual to prepare against a failure.
*2
By changing the parameters of
,
and
, “zero calibration” and “span
calibration” can be adjusted optionally. (Digital span accuracy approximately 1/5000. The
accuracy varies depending on the load cell output accuracy and the conditions of calibration.)
Except for an emergency, perform calibration with an actual load.
9.7500
to 9.8500
Gravity acceleration
of place of calibration
Gravity acceleration of the place of calibration.
Unit
:
in m/s
2
.
9.8000
9.7500
to 9.8500
Gravity acceleration
of place of use
Gravity acceleration of the place where the scale is
being used. Unit
:
in m/s
2
.
9.8000
0 to 1
Disable hold
0
:
Enables.
1
:
Disables.
0
Input voltage
Displayed weight
0