- 46 -
AD-4430B
Function No.
Parameter
and range
Item name
Description
Default
value
User
setting MEMO
0 to 1
Output when
overflow and
unstable
Standard serial output when the weight value
overflows and is unstable.
0
:
Disables output.
1
:
Enables output.
1
1 to 3
Exceeding
negative
gross weight
To judge when the negative gross weight is
exceeded.
1
:
Gross weight < -99999
2
:
Gross weight < Negative weighing capacity
3
:
Gross weight < -19d
1
1 to 2
Exceeding
negative net
weight
To judge when the negative net weight is
exceeded.
1
:
Net weight < -99999
2
:
Net weight < Negative weighing capacity
1
0 to 1
Clear the zero
value
Select whether or not to clear the zero value.
0
:
Disables.
1
:
Enables.
1
0 to 1
Zero setting
when power
is turned on
Select whether or not to perform zero setting
when power is turned on.
0
:
Disables.
1
:
Enables.
0
-7.0000
to 7.0000
Input voltage
at zero
Input voltage from a load cell at zero. Unit
:
mV/V.
This value is determined in zero calibration
during the calibration with an actual load.
0.0000
0.0001
to 9.9999
Input voltage
at span
Input voltage from a load cell at span. Unit
:
mV/V.
This value and the value of are determined in
span calibration during the calibration with an
actual load.
3.2000
1 to 99999
Calibration
weight
corresponding
to Input voltage
at span
The input voltage at span for
displays
the input voltage per this setting of weighing
value.
When performing digital span, which
calibrates without using a calibration weight,
setting this function so as
,
and
are required.
32000
9.7500
to 9.8500
Gravity
acceleration
of place of
calibration
Gravity acceleration of the place of calibration.
Unit
:
in m/s
2
.
9.8000
9.7500
to 9.8500
Gravity
acceleration of
place of use
Gravity acceleration of the place where the
scale is being used. Unit
:
in m/s
2
.
9.8000
0 to 1 Disable hold
0
:
Enables.
1
:
Disables.
0