- 24 -
AD-4430B
5.3.5.
The Actual Load Linearization Function (
)
Set the digital linearization by loading/unloading masses.
Warm up the module for at least ten minutes to avoid the effects of temperature drift.
The input order should proceed from the smallest mass to the largest mass.
Step 1 Press the
+
key to enter to the calibration mode and
display
. Press the
key to start the calibration
and display
. Select
using the
key
and press the
key.
Step 2
is displayed.
If monitoring the current weighing value, press the
key.
When pressing the
key again,
is display.
Step 3 Placed nothing on the pan and wait for the stabilization ( LED ).
Press the
key.
is displayed for approximately
two seconds.
Step 4
is displayed.
If you want to check the current weighing value, press the
key.
When pressing the
key again,
is displayed.
Press the
key. The weight value (the current weighing
capacity) is displayed and the least digit of the value blinks.
Correct the value using the
and
key so as to be the
weight value used.
Step 5 Place the weight on the pan. Wait for the stabilization ( LED ).
Press the
key.
is displayed for approximately
two seconds.
Step 5
is displayed. Repeat step 4 and step 5. The procedure
proceeds in order of
→
→
.
Step 7 If finishing the input, proceed to step 8.
If you want to re-input the digital linearization, select the input point
using the
key. All data following the new input point will be
cleared.
Step 8 Press the
key.
is displayed and the inputted
data will be stored in the FRAM. At the same time, the calibrated
data is also refreshed. Press again the
key to return to
weighing mode.
If
X
is displayed, an error has occurred. X : error number.
Refer to “
5.3.8. Error Codes For The Calibration
” to take some measures.
The blinking decimal point means that the current value is not the weight value.
Example