![AND AD-4430B Скачать руководство пользователя страница 49](http://html1.mh-extra.com/html/and/ad-4430b/ad-4430b_instruction-manual_2941312049.webp)
AD-4430B
- 47 -
7.6.2.
The Linearization Function
(
)
Function No.
Parameter
and range
Item name
Description
Default
value
User
setting MEMO
0 to 5
Number of
input points
Number of points where linear input was done.
The linear-zero input is included as one point.
Digital linearization is not performed when the
set value is between 1 and 2.
0
-7.0000
to 7.0000
Linear-zero Voltage for linear-zero input. Unit
:
mV/V.
0.0000
0 to 99999
Linear 1
Mass value
The mass value for linear 1 input.
The decimal point position depends on
.
0
0.0000 to
9.9999
Linear 1
Span
voltage
The span voltage between linear-zero and linear
1 input. Unit
:
mV/V.
0.0000
0 to 99999
Linear 2
Mass value
The mass value for linear 2 input.
The decimal point position depends on
.
0
0.0000
to 9.9999
Linear 2
Span
voltage
The span voltage between linear-zero and linear
2 input. Unit
:
mV/V.
0.0000
0 to 99999
Linear 3
Mass value
The mass value for linear 3 input.
The decimal point position depends on
.
0
0.0000
to 9.9999
Linear 3
Span
voltage
The span voltage between linear-zero and linear
3 input. Unit
:
mV/V.
0.0000
0 to 99999
Linear 4
Mass value
The mass value for linear 4 input.
The decimal point position depends on
.
0
0.0000 to
9.9999
Linear 4
Span
voltage
The span voltage between linear-zero and linear
4 input. Unit
:
mV/V.
0.0000
7.6.3.
The Basics Function
(
)
Function No.
Parameter
and range
Item name
Description
Default
value
User
setting MEMO
0000 to 1111
key switch
disable
Each digit of the setting corresponds to a key
switch. Only available in the weighing mode.
0
:
Enables Key assignment to each binary digit.
1
:
Disables 4th
3rd 2nd
1st
0000
(binary)
0 to 7
key
function
0
:
None
1
:
Manual print command
2
:
Hold
3
:
Alternate switch
4
:
Momentary switch
5
:
Changing between gross weight and net weight
6
:
Clear the tare weight
7
:
Clear the zero value
5