![AND AD-4430B Скачать руководство пользователя страница 28](http://html1.mh-extra.com/html/and/ad-4430b/ad-4430b_instruction-manual_2941312028.webp)
- 26 -
AD-4430B
Function No.
Parameter
and range
Item name
Description
Default
value
0 to 0.0000
Decimal point
position
Decimal point position of the weight value
0
0.000
0.0
0.0000
0.00
0
1 to 50
Minimum division
Minimum division (d) of the weight value
1
10
2
20
5
50
1
1 to 99999 Weighing capacity
Weighing capacity of the module.
Weighing is possible up to the value of this setting plus
8 d. If the value exceeds this, overflow will occur and
will not be displayed. The decimal point position is the
same as the setting of
.
70000
0 to 100
Zero range
Range to enable zero adjustment by the
(ZERO)
key Expressed as a percentage of the weighing
capacity with the calibration zero point as the center.
For example, if this is set to 2, the value in the range of
±2% of the weighing capacity with the calibration zero
point at the center will be to zero. When a power-ON
zero is performed, the initial zero point will be the
center.
2
0.0 to 5.0
Zero tracking time
Performs zero tracking using this setting in
combination with the setting of the zero tracking width
. When 0.0, zero tracking will not be performed.
Unit
:
Second in 0.1 increments.
0.0
0.0 to 9.9
Zero tracking width
Performs zero tracking using this setting in
combination with the setting of the zero tracking time
. When 0.0, zero tracking will not be performed.
Unit
:
d (minimum division) in 0.1 increments.
0.0
Weight value
Weight value
0 d
4.5 d
When
= 1.0,
= 4.5
1 second
Zero tracking follows the weight
value drifting around the zero point
and adjusts to display as zero.
d = Minimum division = 1 digit
Zero tracking functions when the
weight value is drifting the range
shown in the graph.
1 second
2 second
5.0 d
4.5 d
4.0 d
3.5 d
3.0 d
2.5 d
2.0 d
1.5 d
1.0 d
0.5 d
0.0 d
When
= 2.0,
= 0.5
When
= 1.0,
= 4.5