ANALOX 1371 MK 2 SURVEYOR CARBON DIOXIDE MONITOR
1371-810-03 User Manual, August 2007
617
2
SPECIFICATIONS
2.1
GENERAL
Ranges Available
: 0 to 0.5%,1%, 2%, 5%, 10%, 20%, 100% CO
2
Linearity
: 1% FS
Repeatability
: 1% FS
Resolution
: ± 1 Digit on Display
Zero Drift
: 2% FSD / week max
Response time
: Less than 30 secs to T90
Display
: 3 1/2 Digit LCD (1999 max reading)
Alarm Ranges
: Adjustable over full range of Measurement
2.2
OPERATIONAL
Ambient Temperature
: 5 to 50
°
C (Wider operating Temperature Units are
available on request)
Storage Temperature
: -20 to 70
°
C
Analyser Temp.
: Ambient
Warm-up Time
: 2 mins at 25
°
C (Typical)
Temp. Co-efficient
: Zero : 0.1% FS /
°
C (Typical)
: Span : 0.2% FS /
°
C (Typical)
Humidity
: 0 -90% Non-condensing
Pump Flow Rate
: 75 ml/min (4.5 l/hr)
Pressure Effect
: 0.15% of Reading per millibar
Operating Pressure
: 0.9 to 1.1 Bar ABS.
Power Supplies
: a) 230v AC 50/60Hz
(specified at time of order)
: b) 110v AC 50/60 Hz