ANALOX 1371 MK 2 SURVEYOR CARBON DIOXIDE MONITOR

 

 
 
 

 

 
1371-810-03 User Manual, August 2007 

517

 

INTRODUCTION

 

 
 

The  ANALOX  1371  SURVEYOR  provides  a  clear  Digital  Display  of  Carbon  Dioxide 
concentration  in  a gas  sample.    It  can  be   employed  either   as  an ambient monitor,  using its 
internal pump to  draw  the  sample,  or  as  an  'On-Line'  monitor if the gas to be  checked  is 
flowing through a pipe. 

 

The Instrument uses a 'State-of-the-Art'  Infra-Red Analyser, with  no  electro-mechanical moving 
parts,    designed  to  meet  the  demand    for    accurate  measurements  of  Carbon  Dioxide,  where 
performance and quality are  essential.  

      

Ranges of measurement available are as follows : 

      

Range   

 

Part Number 

0-100% 

 

SA1 D03 BG88 N002 

0-20%   

SA1 D03 BE87 N002 

0 - 10% 

 

SA1 D03 BD86 N002 

0 - 5%   

SA1 D03 BC89 N002 

0 - 2%   

SA1 D03 BB90 N002 

0 - 1%   

SA1 D03 BA91 N002 

0-0.5%  

SA1 D03 SA92 N541 

 
      

Two  completely  independent,  user adjustable alarm  circuits  are incorporated,  the  setpoints  of  
which are accessible on  the  front panel,  and  adjustable  over the full range of the  Instrument.  
Both audible  and  visual indication is provided at the onset of  an  alarm condition and volt-free 
relay contacts are available for connection of external warning devices. 

      

The  Instrument  is  housed  in  a  high  impact,  glass  reinforced polycarbonate  wall  mounting  
case and is sealed to  IP65,  when  the hinged  transparent  front  cover is closed.  This  front  cover 
 also incorporates  a  simple  locking  device  to  discourage  unauthorised  adjustments of the front 
panel  controls.  The  Instrument  case  has  been  coated with  a  double  layer  of conductive  
material  to  provide EMC Compliance -  ie  Immunity  from  external  Radio Frequency Emissions 
and  containment  of  any  internally  generated    RF.    Mains  input  filters  are  also  fitted  to  avoid 
conducted interference. 

      

Installation    is  straightforward  only  requiring  three  fixing  points  and  all  external  electrical 
connections are made to screw terminals. 

      

Содержание 1371 MK2

Страница 1: ...R 1371 810 03 User Manual August 2007 117 ANALOX 1371 MK2 SURVEYOR CARBON DIOXIDE MONITOR INSTRUCTION MANUAL Analox Sensor Techn 15 Ellerbeck Court Stokesley Business Park STOKESLEY Cleveland TS9 5PT...

Страница 2: ...ONING 8 3 1 INSTALLATION 8 3 2 CONNECTIONS 9 3 3 COMMISSIONING 9 4 OPERATION 10 4 1 ALARM SETPOINTS 10 5 CALIBRATION 11 5 1 GENERAL COMMENTS 11 5 2 CALIBRATION PROCEDURE 11 6 MAINTENANCE 14 6 1 GENERA...

Страница 3: ...o be correct at the time of publication The purchasers attention is drawn to our terms and conditions of sale a copy of which is available on request Further technical assistance may be obtained by co...

Страница 4: ...ANALOX 1371 MK 2 SURVEYOR CARBON DIOXIDE MONITOR 1371 810 03 User Manual August 2007 417...

Страница 5: ...ndependent user adjustable alarm circuits are incorporated the setpoints of which are accessible on the front panel and adjustable over the full range of the Instrument Both audible and visual indicat...

Страница 6: ...larm Ranges Adjustable over full range of Measurement 2 2 OPERATIONAL Ambient Temperature 5 to 50 C Wider operating Temperature Units are available on request Storage Temperature 20 to 70 C Analyser T...

Страница 7: ...ensions 296mm W 256mm H 118mm D Weight 3 2 kg Gas Connections Suitable for 6mmOD 4mmID plastic Pipe NOTE DISPLAY RESOLUTION ACCURACY The Digital displays used in this series of instruments are of the...

Страница 8: ...e two screws from the Terminal compartment cover situated along the lower edge of the Instrument Remove the cover taking care not to strain the connecting wires to the audible warning buzzer These wir...

Страница 9: ...the Terminal compartment 3 2 3 On DC Supply versions a 12V DC supply should be connected to the terminals marked 12V and 3 2 4 Remote gas sampling may be carried out by connecting up to about 30 metr...

Страница 10: ...tate The relayand LED will remain in this condition until the measured concentration falls below the alarm setpoint at which time the relay will revert to its normal state and the LED will be turned o...

Страница 11: ...ed this is quite normal Connect the delivery tube to the small Barbed connector on the flowmeter and then to the GAS IN port on the Instrument Adjust the rotary valve on the flowmeter until the indica...

Страница 12: ...045 5 2 5 If during the Calibration process the desired readings cannot be achieved then it is possible that the Infra Red Transducer module requires service To check that the Transducer module is ope...

Страница 13: ...nstrument and open the front cover Remove the four corner screws from the front panel and carefully ease the front panel away from the Instrument case The panel is attached to the main circuit board b...

Страница 14: ...ange of measurement and the gases can be identified by use of narrow band interference filters tuned to an appropriate characteristic absorption wavelength The family of Transducers used in the ANALOX...

Страница 15: ...rn to us would be repaired or replaced without cost or charge but in our discretion we may elect instead to provide to you which ever is the lesser of the cost of replacement or a refund of net purcha...

Страница 16: ...ems must be properly and sufficiently packed 8 2 REPAIR Apart from routine Calibration and filter changes there are no user serviceable parts inside the Instrument In the event of any suspected malfun...

Страница 17: ...f the housing 2 The Transducer is located in the lower right corner of the housing It is a metal box measuring approximately 115mm x 65mm x 30mm 3 Remove the 6mm OD plastic tubes from the gas flow ada...

Страница 18: ...Housing If obviously not operating contact Agent or SAEP Ltd Check any external tubing to the Instrument for blockage kinks No Display response to Calibration Gas Check Calibration Gas Cylinder is not...

Страница 19: ...0 10 FS SA7 D03 10P 20 FS SA7 D03 20P 100 FS SA7 D03 100P 10 2 CALIBRATION SPARES Gases 20 Litre Cylinder Disposable Test Gas Pure Nitrogen SA7 L20 01 0 5 CO2 in Air SA7 L20 130 1 5 CO2 in Air SA7 L20...

Отзывы: