CALIBRATION

 

      

5.1 

GENERAL COMMENTS

 

      

Calibration  requires  gases  to be fed from  cylinders  of  certified concentration.  These gases 
MUST NOT be delivered directly at cylinder pressure  to  the  Instrument,  otherwise the internal  
pump  will  be damaged.  Scottish Anglo E.P. Ltd. supply calibration kits for the 1371 Instrument, 
comprising two cylinders of gas, a cylinder mounted flowmeter with control valve and suitable 
tubing. It is strongly recommended that this kit is used for Calibration. 

 

If  alternative  calibration  equipment  is  used,  precautions  must  be  taken  to  limit  the  gas 
pressure and flowrate to a value not exceeding 1.0 Litres per Minute.

 

 

Under no circumstances must excessive gas pressure or flow be applied to the instrument 
otherwise damage to the pump and/or Transducer will be inevitable. 

 

      

5.2 

CALIBRATION PROCEDURE

 

      
5.2.1  Zero calibration 
 

Using  a  Calibration  Kit  cylinder  of  'Zero  Gas'  ie    Nitrogen,  fit  the  Flowmeter/control  valve 
assembly on to the Cylinder screw connector. There may be a small escape of gas just before the 
assembly is fully tightened - this is quite normal. Connect the delivery tube to the small Barbed 
connector on the flowmeter and then to the 'GAS IN' port on the Instrument.  Adjust the rotary 
valve on the flowmeter until the indicator ball is level with the middle graduation on the flowmeter. 
This produces a flow of approximately 1 Litre per minute.  Allow the displayed reading to stabilise 
and if necessary,  adjust the 'ZERO' trimmer,  to the left of the display, until the Display shows 
'000'. 

      
5.2.2  Span Calibration 
 

Using a Calibration Kit cylinder of 'Span Gas'  having a concentration  of CO

2

 between 75% and 

100% of Instrument measuring  range, fit the Flowmeter/control valve as described in Section 
5.2.1.  Adjust the rotary valve on the flowmeter until the indicator ball is level with the middle 
graduation on the flowmeter. This produces a flow of approximately 1 Litre per minute. Allow the 
displayed reading to stabilise and if necessary,  adjust the  'SPAN'  trimmer, until  the  reading  on 
the display is the same as the concentration of the Certified  gas mixture. 

 
5.2.3  Close  the  control  valve, remove the plastic tubing from the instrument and the flowmeter and 

remove  the  flowmeter/control  valve  assembly  from  the  gas  cylinder.  The  control  valve  is  not 
designed to provide a total gas shut-off when closed so to conserve gas supplies, it should only be 
fitted immediately before use. 

      

Содержание 1371 MK2

Страница 1: ...R 1371 810 03 User Manual August 2007 117 ANALOX 1371 MK2 SURVEYOR CARBON DIOXIDE MONITOR INSTRUCTION MANUAL Analox Sensor Techn 15 Ellerbeck Court Stokesley Business Park STOKESLEY Cleveland TS9 5PT...

Страница 2: ...ONING 8 3 1 INSTALLATION 8 3 2 CONNECTIONS 9 3 3 COMMISSIONING 9 4 OPERATION 10 4 1 ALARM SETPOINTS 10 5 CALIBRATION 11 5 1 GENERAL COMMENTS 11 5 2 CALIBRATION PROCEDURE 11 6 MAINTENANCE 14 6 1 GENERA...

Страница 3: ...o be correct at the time of publication The purchasers attention is drawn to our terms and conditions of sale a copy of which is available on request Further technical assistance may be obtained by co...

Страница 4: ...ANALOX 1371 MK 2 SURVEYOR CARBON DIOXIDE MONITOR 1371 810 03 User Manual August 2007 417...

Страница 5: ...ndependent user adjustable alarm circuits are incorporated the setpoints of which are accessible on the front panel and adjustable over the full range of the Instrument Both audible and visual indicat...

Страница 6: ...larm Ranges Adjustable over full range of Measurement 2 2 OPERATIONAL Ambient Temperature 5 to 50 C Wider operating Temperature Units are available on request Storage Temperature 20 to 70 C Analyser T...

Страница 7: ...ensions 296mm W 256mm H 118mm D Weight 3 2 kg Gas Connections Suitable for 6mmOD 4mmID plastic Pipe NOTE DISPLAY RESOLUTION ACCURACY The Digital displays used in this series of instruments are of the...

Страница 8: ...e two screws from the Terminal compartment cover situated along the lower edge of the Instrument Remove the cover taking care not to strain the connecting wires to the audible warning buzzer These wir...

Страница 9: ...the Terminal compartment 3 2 3 On DC Supply versions a 12V DC supply should be connected to the terminals marked 12V and 3 2 4 Remote gas sampling may be carried out by connecting up to about 30 metr...

Страница 10: ...tate The relayand LED will remain in this condition until the measured concentration falls below the alarm setpoint at which time the relay will revert to its normal state and the LED will be turned o...

Страница 11: ...ed this is quite normal Connect the delivery tube to the small Barbed connector on the flowmeter and then to the GAS IN port on the Instrument Adjust the rotary valve on the flowmeter until the indica...

Страница 12: ...045 5 2 5 If during the Calibration process the desired readings cannot be achieved then it is possible that the Infra Red Transducer module requires service To check that the Transducer module is ope...

Страница 13: ...nstrument and open the front cover Remove the four corner screws from the front panel and carefully ease the front panel away from the Instrument case The panel is attached to the main circuit board b...

Страница 14: ...ange of measurement and the gases can be identified by use of narrow band interference filters tuned to an appropriate characteristic absorption wavelength The family of Transducers used in the ANALOX...

Страница 15: ...rn to us would be repaired or replaced without cost or charge but in our discretion we may elect instead to provide to you which ever is the lesser of the cost of replacement or a refund of net purcha...

Страница 16: ...ems must be properly and sufficiently packed 8 2 REPAIR Apart from routine Calibration and filter changes there are no user serviceable parts inside the Instrument In the event of any suspected malfun...

Страница 17: ...f the housing 2 The Transducer is located in the lower right corner of the housing It is a metal box measuring approximately 115mm x 65mm x 30mm 3 Remove the 6mm OD plastic tubes from the gas flow ada...

Страница 18: ...Housing If obviously not operating contact Agent or SAEP Ltd Check any external tubing to the Instrument for blockage kinks No Display response to Calibration Gas Check Calibration Gas Cylinder is not...

Страница 19: ...0 10 FS SA7 D03 10P 20 FS SA7 D03 20P 100 FS SA7 D03 100P 10 2 CALIBRATION SPARES Gases 20 Litre Cylinder Disposable Test Gas Pure Nitrogen SA7 L20 01 0 5 CO2 in Air SA7 L20 130 1 5 CO2 in Air SA7 L20...

Отзывы: