ANALOX 1371 MK 2 SURVEYOR CARBON DIOXIDE MONITOR
1371-810-03 User Manual, August 2007
817
3
INSTALLATION AND COMMISSIONING
3.1
INSTALLATION
3.1.1 The 1371 Surveyor should be mounted on a vertical surface using 3 off M4 machine screws or No.
8 woodscrews. Make sure that the chosen mounting position allows sufficient room on the Left
of the Instrument, to permit the hinged front cover to be fully opened, and below the unit to
allow cables and pipework to be correctly fitted. The Hinged front cover may be reversed if
required.
3.1.2 Remove the two screws from the Terminal compartment cover, situated along the lower edge
of the Instrument. Remove the cover taking care not to strain the connecting wires to the audible
warning buzzer. These wires are connected to the Motherboard using a small 2 way plug. This
may be removed by carefully releasing the locking latch on the Left of the plug.
3.1.3 Referring to the Figure 1 below, drill three holes in the mounting surface or fit 3 woodscrews at
the hole centres.
3.1.4 Fix the Top screw in place but not fully screwed down - this is required to hang the Instrument
from the Keyhole position on the back of the case. Hang the Instrument on the central screw and
then secure it with two screws in the slotted positions in the lower corners.