background image

 

ANALOX 1371 MK 2 SURVEYOR CARBON DIOXIDE MONITOR

 

 
 
 

 

 
1371-810-03 User Manual, August 2007 

917

 

 

Figure 1. Fixing Centres. 

 

3.2  

CONNECTIONS 

 

      
3.2.1  Connections  to remote equipment,  eg Flashing beacons or  High powered Sounders and data 

recording devices are made through two cable glands,  to  the  screw terminals in the middle area of 
 the  terminal compartment.  The  upper  row of terminals provide connections to  the alarm  relay  
contacts and the lower row are signal input  and  output connections.  On the 1371 Instrument,  
only the extreme right hand two  terminals  are  used on the Signal terminals  - these provide a 4-
20mA  DC  output  signal, representing  0 to 100% of Instrument measuring range. Optionally, a 0 - 
1volt output can be provided instead of the 4-20mA, if this is requested when ordering. 

      
3.2.2  On  AC  Supply  versions,  an  AC  mains  supply  of  correct  value  as  stated  on  the  Internal  label  

covering  the terminals,  should be connected through the  left most cable gland,  to the terminals 
marked 'L' and 'N'. The Earth lead should  be fixed to the Ring connector provided on the Earthing  
stud, on the Left inside wall of the Terminal compartment. 

 
3.2.3  On DC Supply versions, a 12V DC supply should be connected to the terminals marked '12V', '+' 

and '-'. 

 
3.2.4  Remote gas sampling may be carried out by connecting up to about  30  metres of 4mm I.D.  6mm 

OD plastic tubing between the  'GAS IN'  bulkhead connector and the area to be monitored. 

 
3.2.5  A Pump failure or gas flow failure Alarm circuit is incorporated and this operates a relay, whose 

contacts are connected to a 3way screw terminal block, just to the left of the terminals described in 
Section 3.2.1 Under normal flow conditions, the NC and CM contacts are closed and the NO and 
CM contacts open. Under flow fail or power fail conditions, these contacts will change over. 

      
 

3.3  

COMMISSIONING

 

 
3.3.1  When the 1371 Surveyor is first switched on, it requires a period of about 40 Seconds, for the IR 

Transducer to settle. During this period, the Reading on the Display will be clamped to read the 
Full Scale Value of the particular Instrument. 

 
3.3.2   When the settling time has expired, the Instrument will display the CO

2

 concentration in the gas 

being drawn through the Transducer, by the sample pump. 

 
3.3.4  The  Alarm  setpoints  should be checked  and  adjusted  if necessary,  following  the procedure 

outlined in Section 4 and a  full Calibration carried out as detailed in Section 5. 

      

Содержание 1371 MK2

Страница 1: ...R 1371 810 03 User Manual August 2007 117 ANALOX 1371 MK2 SURVEYOR CARBON DIOXIDE MONITOR INSTRUCTION MANUAL Analox Sensor Techn 15 Ellerbeck Court Stokesley Business Park STOKESLEY Cleveland TS9 5PT...

Страница 2: ...ONING 8 3 1 INSTALLATION 8 3 2 CONNECTIONS 9 3 3 COMMISSIONING 9 4 OPERATION 10 4 1 ALARM SETPOINTS 10 5 CALIBRATION 11 5 1 GENERAL COMMENTS 11 5 2 CALIBRATION PROCEDURE 11 6 MAINTENANCE 14 6 1 GENERA...

Страница 3: ...o be correct at the time of publication The purchasers attention is drawn to our terms and conditions of sale a copy of which is available on request Further technical assistance may be obtained by co...

Страница 4: ...ANALOX 1371 MK 2 SURVEYOR CARBON DIOXIDE MONITOR 1371 810 03 User Manual August 2007 417...

Страница 5: ...ndependent user adjustable alarm circuits are incorporated the setpoints of which are accessible on the front panel and adjustable over the full range of the Instrument Both audible and visual indicat...

Страница 6: ...larm Ranges Adjustable over full range of Measurement 2 2 OPERATIONAL Ambient Temperature 5 to 50 C Wider operating Temperature Units are available on request Storage Temperature 20 to 70 C Analyser T...

Страница 7: ...ensions 296mm W 256mm H 118mm D Weight 3 2 kg Gas Connections Suitable for 6mmOD 4mmID plastic Pipe NOTE DISPLAY RESOLUTION ACCURACY The Digital displays used in this series of instruments are of the...

Страница 8: ...e two screws from the Terminal compartment cover situated along the lower edge of the Instrument Remove the cover taking care not to strain the connecting wires to the audible warning buzzer These wir...

Страница 9: ...the Terminal compartment 3 2 3 On DC Supply versions a 12V DC supply should be connected to the terminals marked 12V and 3 2 4 Remote gas sampling may be carried out by connecting up to about 30 metr...

Страница 10: ...tate The relayand LED will remain in this condition until the measured concentration falls below the alarm setpoint at which time the relay will revert to its normal state and the LED will be turned o...

Страница 11: ...ed this is quite normal Connect the delivery tube to the small Barbed connector on the flowmeter and then to the GAS IN port on the Instrument Adjust the rotary valve on the flowmeter until the indica...

Страница 12: ...045 5 2 5 If during the Calibration process the desired readings cannot be achieved then it is possible that the Infra Red Transducer module requires service To check that the Transducer module is ope...

Страница 13: ...nstrument and open the front cover Remove the four corner screws from the front panel and carefully ease the front panel away from the Instrument case The panel is attached to the main circuit board b...

Страница 14: ...ange of measurement and the gases can be identified by use of narrow band interference filters tuned to an appropriate characteristic absorption wavelength The family of Transducers used in the ANALOX...

Страница 15: ...rn to us would be repaired or replaced without cost or charge but in our discretion we may elect instead to provide to you which ever is the lesser of the cost of replacement or a refund of net purcha...

Страница 16: ...ems must be properly and sufficiently packed 8 2 REPAIR Apart from routine Calibration and filter changes there are no user serviceable parts inside the Instrument In the event of any suspected malfun...

Страница 17: ...f the housing 2 The Transducer is located in the lower right corner of the housing It is a metal box measuring approximately 115mm x 65mm x 30mm 3 Remove the 6mm OD plastic tubes from the gas flow ada...

Страница 18: ...Housing If obviously not operating contact Agent or SAEP Ltd Check any external tubing to the Instrument for blockage kinks No Display response to Calibration Gas Check Calibration Gas Cylinder is not...

Страница 19: ...0 10 FS SA7 D03 10P 20 FS SA7 D03 20P 100 FS SA7 D03 100P 10 2 CALIBRATION SPARES Gases 20 Litre Cylinder Disposable Test Gas Pure Nitrogen SA7 L20 01 0 5 CO2 in Air SA7 L20 130 1 5 CO2 in Air SA7 L20...

Отзывы: