background image

 

ANALOX 1371 MK 2 SURVEYOR CARBON DIOXIDE MONITOR

 

 
 
 

 

 
1371-810-03 User Manual, August 2007 

1017

 

OPERATION

 

      

4.1 

ALARM SETPOINTS

 

      
4.1.1  The  Instrument incorporates two completely  independent  alarm  circuits,  which may be adjusted 

by  the  user,    to  any  point  within  the  full    range  of  the  instrument.    Both  alarm  circuits  are 
configured  to operate  when the measured gas concentration Exceeds the setpoint - ie they are High 
going alarms. 

      
4.1.2  When an alarm condition occurs,  the appropriate Red LED (HI or LO)  will start to flash,  the 

internal Audible device will sound  and the  associated  relay  will change state.  The Audible alarm 
 may  be silenced by pressing the 'MUTE'  button,  located just under the lower edge of the main 
panel. This action also changes the LED from flashing to  a steady illuminated state.  The relay and 
LED will remain in this condition,  until  the  measured concentration falls below  the  alarm  
setpoint, at which time, the relay will revert to its normal state and the LED will be turned off. 

      
4.1.3  Adjustment of the setpoints is a simple procedure.  Locate the switch  marked 'HI - LO'  in the 

'ALARM ADJUST'  section of the  front panel. This device is normally spring biassed to its central 
position. To read the present setpoint for the LO alarm, press the switch to the right - the setpoint is 
then shown on the display. To read the present setpoint for the HI alarm, press the switch to the left 
- the setpoint is then shown on the display. To adjust the setpoint, hold the switch in the appropriate 
position  and  using  the  trimming  tool  provided,  or  a  small  terminal  screwdriver,  rotate    the  
appropriate  trimmer  ,    'HI'    or  'LO'    until    the    desired  setpoint  is  indicated  on  the  Display. 
Clockwise rotation increases the setpoint  value.  Release the switch and repeat for the other alarm 
if required. 

Содержание 1371 MK2

Страница 1: ...R 1371 810 03 User Manual August 2007 117 ANALOX 1371 MK2 SURVEYOR CARBON DIOXIDE MONITOR INSTRUCTION MANUAL Analox Sensor Techn 15 Ellerbeck Court Stokesley Business Park STOKESLEY Cleveland TS9 5PT...

Страница 2: ...ONING 8 3 1 INSTALLATION 8 3 2 CONNECTIONS 9 3 3 COMMISSIONING 9 4 OPERATION 10 4 1 ALARM SETPOINTS 10 5 CALIBRATION 11 5 1 GENERAL COMMENTS 11 5 2 CALIBRATION PROCEDURE 11 6 MAINTENANCE 14 6 1 GENERA...

Страница 3: ...o be correct at the time of publication The purchasers attention is drawn to our terms and conditions of sale a copy of which is available on request Further technical assistance may be obtained by co...

Страница 4: ...ANALOX 1371 MK 2 SURVEYOR CARBON DIOXIDE MONITOR 1371 810 03 User Manual August 2007 417...

Страница 5: ...ndependent user adjustable alarm circuits are incorporated the setpoints of which are accessible on the front panel and adjustable over the full range of the Instrument Both audible and visual indicat...

Страница 6: ...larm Ranges Adjustable over full range of Measurement 2 2 OPERATIONAL Ambient Temperature 5 to 50 C Wider operating Temperature Units are available on request Storage Temperature 20 to 70 C Analyser T...

Страница 7: ...ensions 296mm W 256mm H 118mm D Weight 3 2 kg Gas Connections Suitable for 6mmOD 4mmID plastic Pipe NOTE DISPLAY RESOLUTION ACCURACY The Digital displays used in this series of instruments are of the...

Страница 8: ...e two screws from the Terminal compartment cover situated along the lower edge of the Instrument Remove the cover taking care not to strain the connecting wires to the audible warning buzzer These wir...

Страница 9: ...the Terminal compartment 3 2 3 On DC Supply versions a 12V DC supply should be connected to the terminals marked 12V and 3 2 4 Remote gas sampling may be carried out by connecting up to about 30 metr...

Страница 10: ...tate The relayand LED will remain in this condition until the measured concentration falls below the alarm setpoint at which time the relay will revert to its normal state and the LED will be turned o...

Страница 11: ...ed this is quite normal Connect the delivery tube to the small Barbed connector on the flowmeter and then to the GAS IN port on the Instrument Adjust the rotary valve on the flowmeter until the indica...

Страница 12: ...045 5 2 5 If during the Calibration process the desired readings cannot be achieved then it is possible that the Infra Red Transducer module requires service To check that the Transducer module is ope...

Страница 13: ...nstrument and open the front cover Remove the four corner screws from the front panel and carefully ease the front panel away from the Instrument case The panel is attached to the main circuit board b...

Страница 14: ...ange of measurement and the gases can be identified by use of narrow band interference filters tuned to an appropriate characteristic absorption wavelength The family of Transducers used in the ANALOX...

Страница 15: ...rn to us would be repaired or replaced without cost or charge but in our discretion we may elect instead to provide to you which ever is the lesser of the cost of replacement or a refund of net purcha...

Страница 16: ...ems must be properly and sufficiently packed 8 2 REPAIR Apart from routine Calibration and filter changes there are no user serviceable parts inside the Instrument In the event of any suspected malfun...

Страница 17: ...f the housing 2 The Transducer is located in the lower right corner of the housing It is a metal box measuring approximately 115mm x 65mm x 30mm 3 Remove the 6mm OD plastic tubes from the gas flow ada...

Страница 18: ...Housing If obviously not operating contact Agent or SAEP Ltd Check any external tubing to the Instrument for blockage kinks No Display response to Calibration Gas Check Calibration Gas Cylinder is not...

Страница 19: ...0 10 FS SA7 D03 10P 20 FS SA7 D03 20P 100 FS SA7 D03 100P 10 2 CALIBRATION SPARES Gases 20 Litre Cylinder Disposable Test Gas Pure Nitrogen SA7 L20 01 0 5 CO2 in Air SA7 L20 130 1 5 CO2 in Air SA7 L20...

Отзывы: