Amico Beds Corporation
Display
Interpretation
Solution
Load Cell Interface Not Calibrated
This occurs when the Load Cell
Interface has not been calibrated.
• Contact Amico Beds Customer Service for support.
Load Cell Interface Incorrectly Calibrated
This occurs when the Load Cell
Interface requires re-calibration.
• Contact Amico Beds Customer Service for support.
Load Cell #1 is Disconnected
This occurs if Load Cell #1 is
disconnected or has failed.
• Unplug the bed from Mains Power, wait 10 seconds,
then plug the bed back in and check if the error is
cleared.
• If the error code reappears, contact Amico Beds
Customer Service for support.
Reference Signal for Load Cell #1 is
Missing
This occurs if the other load cells
cannot find the reference signal from
Load Cell #1.
• Unplug the bed from Mains Power, wait 10 seconds,
then plug the bed back in and check if the error is
cleared.
• If the error code reappears, contact Amico Beds
Customer Service for support.
Load Cell #2 is Disconnected
This occurs if Load Cell #2 is
disconnected or has failed.
• Unplug the bed from Mains Power, wait 10 seconds,
then plug the bed back in and check if the error is
cleared.
• If the error code reappears, contact Amico Beds
Customer Service for support.
Load Cell #3 is Disconnected
This occurs if Load Cell #3 is
disconnected or has failed.
• Unplug the bed from Mains Power, wait 10 seconds,
then plug the bed back in and check if the error is
cleared.
• If the error code reappears, contact Amico Beds
Customer Service for support.
Load Cell #4 is Disconnected
This occurs if Load Cell #4 is
disconnected or has failed.
• Unplug the bed from Mains Power, wait 10 seconds,
then plug the bed back in and check if the error is
cleared.
• If the error code reappears, contact Amico Beds
Customer Service for support.
NOTE:
All icons and error codes will disappear after 20 seconds, but it does not mean the error is resolved. To make the
code reappear on the display, press any button on the Scale System Controls.
Troubleshooting