
Versitron Operation Manual
59
Click the
Apply
tab will prompt the
Capstan Wheel Adjustment
screen, adjust the
capstan wheel to the optimum load position and when ready click
Done
.
The
Add a New Reading
screen is now ready for the first measurement to be
executed. Place the corresponding test block on the anvil and proceed to take the first
measurement by operating the arm lever or foot pedal, click
OK
if satisfy with the
reading or
Cancel
to retake the measurement.
When the first measurement is completed, the
Indirect Calibration
screen will appear
with the data grid listing the Reference value, the actual Reading and the Scale used.