Versitron Operation Manual
33
5.15
Testing Without Anvil Stage, AT130T Stands
The anvil stage with elevating screw may be removed to increase the test stand
capacity to its maximum.
1.
Snap in clamping shield to protect indenter.
2.
Lower anvil about 2” (5 cm) from indenter.
3.
Loosen anvil plate screw. If necessary, use a spanner wrench in kit
4.
Grip the anvil stage with both hands and lift up and out diagonally.
5.
The tester base can be used directly as an anvil to support large test pieces.
This test base can also be used to support fixtures for testing irregularly
shaped parts.
6.
On large work pieces, use the clamping shield to secure parts. Otherwise,
maintain a distance between indenter and work piece at approximately 1/16”.
Maximum vertical capacity without anvil stage is 16-1/2” (420 mm) (standard
unit). Maximum horizontal reach is 8-1/2” (215 mm). Other T-stand vertical
capacities are available up to 36”. Use the clamping shield on larger openings
to reduce excessive deflection.
7.
When replacing anvil stage, be sure the test plate and bottom supports of the
stage are clean. Then reverse procedures and hand-tighten anvil plate
screw. Do not use the wrench for tightening.
Anvil Plate Screw
Anvil Plate