User Manual | 19
The following defi nes each fi eld:
DDDD-DD-DD TT:TT:TT,#,R,S,A,EEEEEEEE,NNNNN,C, …
DDDD-DD-DD
= Date (2000-03-23)
TT:TT:TT
= Time (12:23:44)
#
= Analyzer Serial Number
R
= Run or Maintenance (Calibration) Mode
A
= EDL or Single Analysis Mode
S
= N/A
EEEEEEEE
= Eight digits of error status (0 = Not Error, 1 = Er-
ror)
1.
General Alarm (MSD)
2. Spare
3. Spare
4. Spare
5. Spare
6. Spare
7. Spare
8.
Spare (LSD)
NNNNN
= Name of impurity
C
= Concentration in ppb x 100
The following is an example of this output string:
2001-08-16 14:35:00,99,R,S,1,10000000,H2 ,0,CO ,2483, , , , , ,
Com Ports - COM 1, COM 2, and COM 3
All three of these ports are 9-pin male D-type connectors. Pin 2 is used
to transmit data to other equipment. The analyzer uses Pin 3 to receive
data. Pin 5 is a common ground. All other pins are unused.
Field 1
Field 2
Field 3
Field 4
Field 5
Field 6
Field 7
Field 8
Date & Time
of RUN
Analyzer
Serial #
Run /
Maintenance
EDL / Single
Analysis
N/A
ta7000
Error Status
Impurity
Name
Concentration
in ppb x 100