18 | Model ta7000F Gas Purity Monitor
Nitrogen Carrier Gas (for Oxygen and Hydrogen models only)
Inlet Fitting
1/4-inch Male face seal fitting*
Inlet pressure range
70 to 90 psig (4.8 to 6.1 bar)
Inlet pressure stability
+ 2%, UHP regulator required
Flow
rate
300
cc/min
Purity
99.9999% (< 1 ppm total impurity co
between
calibration
Chassis
Dimensions
7” (18cm) H x 16.8” (43cm) W x 26.5”
(67cm)
D
Weight
35
lb.
(15.9
kg)
Power
100 - 120 VAC, 50/60 Hz,; 200 - 240
VAC,
50/60
Hz
Outputs
Display
LCD graphics, backlit, 100mm x 150mm
Printer Port
Concentration, chromatogram, and diagnos
tic
reports
Serial PLC Port
Concentration data and general alarm via
RS232
4 to 20ma (optional)
Concentration signals
ta7000 PLC Output Format
The concentration data, status, and other detail information is sent from
port 2 of the ta7000 automatically at the end of every run. The format
used is called Comma Separated Variable or CSV. This format is one of the
common MS EXCEL formats.
The fields within each record are defined in Table 1. Fields 1 – 6 are the
record identification fields and fields 7 & 8 are the impurity fields.
The next impurities will be at fields 9 – 10, 11 – 12, 13 – 14, 15 –16, and 17
– 18. This string is terminated by a CR, carriage return, and LF, line feed,
characters.
The parameters are 9,600 baud, 8 data bits, 1 stop bit and no parity.