User Manual | 17
Actuator Gas
Inlet Fitting
1/8-inch VICI compression fitting
Inlet Pressure Range
70 to 150 psig (4.8 to 10.2 bar)
Gas Type
Inerts (Nitrogen, Argon, Helium), 7
nines or better or CDA, dewpoint
<
-65 °C
Gas Type (for ta7000F-O
2
Inerts (Nitrogen, Argon, Helium), 7
H
2
Only)
nines or better
Calibration Gas
Inlet Fitting
1/16-inch VICI compression fitting
Cylinder concentration
5 to 10 ppm H
2
, CO, CH
4
CO
2
and
ethane (C
2
H
6
) balance nitrogen
Cylinder volume
100 cubic feet minimum (2800L)
Supply pressure
80 to 90 psig (5.3 to 6.1 bar)
Consumption
Approximately 1.0 L per calibration
(4 to 8 L/ month), automatic shut off
FID Hydrogen
Inlet Fitting
1/16-inch VICI compression fitting
Inlet pressure range
50 to 60 psig (3.4 to 4.1 bar)
Inlet pressure stability
+ 2%, UHP regulator required
Flow
rate
35
cc/min
Purity
Hydrocarbons, CO, CO , Moisture <
1 ppm
FID Air
Inlet Fitting
1/16-inch VICI compression fitting
Inlet pressure range
50 to 60 psig (3.4 to 4.1 bar)
Inlet pressure stability
+ 2%, regulator required
Flow
rate
250
cc/min
Purity
Total hydrocarbons < 1 ppm, dew
point
<
-
65°
C