16 | Model ta7000F Gas Purity Monitor
Resolution, Communication Ports
0.01 ppb
Sample Gas Supplies
Inlet Pressure Range
70 to 90 psig (4.8 to 6.1 bar)
Inlet Pressure Stability
± 2%, regulator required
Return Pressure
Atmospheric vent is optimal, ± 0.5 psig
maximum
Flow Rate
300 cc/min minimum, sample bypass at 50
cc/min.
Temperature
60 to 100 °F (16 to 38 °C), optimum when
maintained
±
2°C
Maximum Impurity Levels 200 ppb Moisture, CO
2
, Hydrocarbons
Gas Ports
Carrier/Sample Inlet
1/4-inch VCR compatible
Sample Inlet 1/4-inch VCR compatible (H
2
and O
2
Models only)
FID
Air
1/16-inch
VICI
compression
FID H
2
1/16-inch
VICI
compression
Aux (Calibration)
1/16-inch VICI compression
Sample Outlet
1/16-inch VICI compression
FID H2 SHUT-OFF 1/8-inch VICI compression
(Supplied at 57 psi (3.9 bar))
Actuator Gas
1/8-inch VICI compression
Sample Gas
Inlet Fitting
1/4-inch Male face seal fitting*
Inlet Pressure Range
70 to 90 psig (4.8 to 6.1 bar)
Inlet pressure stability
± 2%, UHP regulator required
Vent pressure
Atmospheric pressure vent is opti
mal,
±
0.5
psig
maximum
FlowRate
300 cc/min minimum, sample bypass
at
50
cc/min
* 1/4-inch face seal x 1/16-inch adapter fi ttings are available from Ametek