User Manual | 39
Calibration Page
Pressing the Page key three times will display the calibration informa-
tion page.
The calibration screen shows information about the most recent cali-
bration run.
Name
is the gas species to be measured.
Conc
is the cylinder concentration (or blend) of the calibration
gas connected to the instrument.
Area
is the area of the last analysis run for each gas type.
Rfactor
is the response factor created by the most recent (or
previous) calibration cycle. Rfactor is used to generate the area.
RT.
is retention time. This is the time point of the peak from
the start of the most recent analysis run.
Conc Factor
is a figure that is preset at the factory. This number
represents the internal gas blender ratio.
Calibration
Screen
Auto Calibration
The Calibration screen is also used to setup automatic calibrations.
The analyzer does not need zero calibration. An external calibration
gas source must be connected to the appropriate port on the rear
panel of the analyzer marked AUX. Calibration gas specifications are
located in the specifications pages in the front of this manual.
Calibration
Idle Mode
12:34:25
Edit Page
Name Conc. Area Rfactor RT
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
CH
4
CO
2
NMHC
4800
42842
5300
38236
10000
40571
*
Conc. Factor 200
Cal Time 00:00
Cal Purge 1 Min
Interval None
Cal Tol 10%