Site Preparation Instructions
AQUATRAN 3/38 Operator’s Manual
A-4
Revision G
MOCON, Inc.
2.
Machinery and air conditioners can generate a large amount of electrical interference on the AC power
lines. These disturbances can interfere with the operation of the system unless steps are taken to isolate
the power disturbances from the power lines serving the instrument. Electrical noise can sometimes be
eliminated by repair, replacement, relocation or electrical filtering of the originating device. If not, a
suitable power line conditioner may have to be installed in the AC line to the system by a qualified
electrician. A noisy circuit or improper grounding will cause adverse performance and possible damage to
the system.
WARNING!
The installation of the appropriate electrical circuit and receptacle requires the skill and
knowledge of a qualified electrician. Do not attempt to test the electrical current in your facility
or to wire or fabricate the electrical power circuit without the services of a qualified electrician.
The MOCON AQUATRAN 3/38 system will provide years of useful life provided these requirements are observed.
Please contact MOCON in the USA at (763) 493-6370 with any questions regarding these Site Preparation
Instructions.
MOCON and AQUATRAN are registered trademarks of MOCON, Inc.