TITLE: Procedure, 5270 Configuration for 4207D
Instrument Model: Model 4207D Compressive Strength Tester
Select the Advanced option.
Change the Setpoint 1 and Setpoint 2 values to 2:133f as shown. This action directs 5270 to download a 0.00
value to the display when a test is started. The “0.00 Lbf” value is already defined in the test profiles for this
instrument in the Start Sequence section of the Test Profile.
Select Tools – Configure – Test Profiles. Select the 4207D instrument and select one of the profiles (note that the
default profiles may not be changed, a copy must be made). Select the Start Sequence option and verify that the
“Download setpoints when test is started” option is enabled, the Signal is set to Load – Maximum and the Value is
0.00 Lbf.