CHANDLER
ENGINEERING
Part Number: 89-0175
Page 1 of 7
TITLE: Procedure, 5270 Configuration for 4207D
Instrument Model: Model 4207D Compressive Strength Tester
Revision
Date
Revised By
Description
Checked By
B 10/01/09 AMH
ECN
T2532
TC
The Model 4207D Compressive Strength Tester may be used with the Model 5270 Data Acquisition and Control
System (DACS). The following procedure is used to configure 5270 for use with this instrument. The following
features are available:
•
Acquisition of peak value from the load display
•
Acquisition of the programmed value from the controller
•
Download of API Fast and Slow load rates to the controller
•
Automatic reset of the load display peak value when a new 5270 test is started
NOTE
: The latest version of the 5270 DACS software must be used for all features to be supported. At this
writing, the most recent version of 5270 is 1.10.097. If an upgrade is needed, please contact AMETEK Chandler
Engineering at 918.250.7200
Using Tools – Configure – I/O Connections, configure an I/O connection using the Modbus protocol.
Select the COM port that is connected to the 4207D instrument. COM1 is illustrated; however, the actual port
assignment may vary as required by the hardware configuration.
Содержание Chandler Engineering 4207D
Страница 8: ...P 4 PREFACE This page is intentionally left blank ...
Страница 18: ...2 8 SECTION 2 OPERATING INSTRUCTIONS This page is intentionally left blank ...
Страница 24: ...3 6 SECTION 3 MAINTENANCE This page is intentionally left blank ...
Страница 28: ...5 2 SECTION 5 REPLACEMENT PARTS This page is intentionally left blank ...
Страница 30: ...6 2 SECTION 6 DRAWINGS AND SCHEMATICS This page is intentionally left blank ...
Страница 39: ...CHANDLER ENGINEERING ...
Страница 40: ......