SECTION 3 – MAINTENANCE 3-1
Section 3 – Maintenance
Maintenance Schedule
MAINTENANCE SCHEDULE
Model 4207D Digital Compressive Strength Tester
COMPONENT
EACH TEST
MONTHLY
3 MONTHS
6 MONTHS
ANNUAL
Oil
Check
Level
Every
two
years,
change the oil
System
Pressure
Check
the
system
pressure, adjust
to 2600 psi
Press Assembly
(See procedure
below)
Perform
maintenance
checks below
Hydraulic Power
Unit (See
procedure below)
Perform
maintenance
checks below
High Temperature
Limiting Circuit
Verify
operation
Platen Control
Switch
Adjust as needed
Load
Cell
Calibrate
This maintenance schedule applies to normal usage of two tests per day. Detailed procedures for these operations are
contained in your manual.
Filling the Oil Reservoir
The system reservoir will not require refilling unless the oil is
drained or a leak occurs. The oil should be changed every two
years. To fill the reservoir, use the following procedure:
1.
Remove the reservoir fill cap.
2.
Fill the reservoir using
10W30
synthetic oil. Verify that
the oil is not contaminated.
Do not reuse old oil.
3.
Fill the reservoir until the oil level is mid-range on the
sight gauge on the front of the reservoir.
Adjusting the System Pressure
The system pressure must be set to a value greater than the maximum pressure required by
the system to create the anticipated loads. Use the following procedure to adjust the system
pressure:
Gauge
Reservoir fill cap
Содержание Chandler Engineering 4207D
Страница 8: ...P 4 PREFACE This page is intentionally left blank ...
Страница 18: ...2 8 SECTION 2 OPERATING INSTRUCTIONS This page is intentionally left blank ...
Страница 24: ...3 6 SECTION 3 MAINTENANCE This page is intentionally left blank ...
Страница 28: ...5 2 SECTION 5 REPLACEMENT PARTS This page is intentionally left blank ...
Страница 30: ...6 2 SECTION 6 DRAWINGS AND SCHEMATICS This page is intentionally left blank ...
Страница 39: ...CHANDLER ENGINEERING ...
Страница 40: ......