3-2
SECTION 3 – MAINTENANCE
Note: The system pressure will decrease as the oil temperature increases. For this
reason, adjust the system pressure when the oil temperature is at or near room
temperature.
1.
Verify that the system temperature is below 80
°
F (27
°
C).
2.
Turn the system
On
.
3.
Place the cylinder control switch in the
Up
position.
4.
Locate the relief valve adjustment and the system pressure
gauge.
5.
Adjust the relief valve until the system pressure gauge
equals 2,600
±
200 psig.
6.
Place the cylinder control switch in the
Off
position.
7.
Tighten the relief valve lock nut.
Resetting the Maximum Temperature Limit
This system is equipped with a circuit that disables the hydraulic power unit if the oil
temperature exceeds 70
°
C. The circuit must be manually reset for continued operation of the
system.
To reset the temperature limit, press the red reset button located on the right panel.
Use the following procedure to adjust the maximum temperature limit:
1.
Turn the system
Off
.
2.
Remove the back panel from the power unit.
3.
Locate the temperature limit circuit board located on the left side of the enclosure
(viewed from the rear).
4.
Locate the calibrated dial on the circuit board. Adjust the knob to a value of 70
°
C.
5.
To reset the temperature circuit, press the
Red
reset button located on the right panel.
6.
To test the circuit, turn the system
On
then adjust the temperature set point below the oil
temperature. Verify that the relay opens and an
LED
illuminates. Adjust the set point
and press the
Reset
button.
Relief Valve Adjustment
Gauge
Содержание Chandler Engineering 4207D
Страница 8: ...P 4 PREFACE This page is intentionally left blank ...
Страница 18: ...2 8 SECTION 2 OPERATING INSTRUCTIONS This page is intentionally left blank ...
Страница 24: ...3 6 SECTION 3 MAINTENANCE This page is intentionally left blank ...
Страница 28: ...5 2 SECTION 5 REPLACEMENT PARTS This page is intentionally left blank ...
Страница 30: ...6 2 SECTION 6 DRAWINGS AND SCHEMATICS This page is intentionally left blank ...
Страница 39: ...CHANDLER ENGINEERING ...
Страница 40: ......