Installation, Operation and Maintenance

Optical Oxygen Analyzer

Additional Options for Product Model Selection (Table 2)

*  Sensor Style - Selection “X” can be: Sxyz, where xyz is equal to fiber length in meters (e.g. S025, S050)

** Mounting Orientation - Selection “Y” can be either “L” or “R”

2�4 

OXYvisor

 Accessories

Table 3 - 

OXYvisor

 Accessories Parts

Part Number

Description

B5008-1125

Pipe Mount Kit - 316 SS (1 to 1.5” pipe)  - Included with OXYvisor (standard)

B4951-1140

Plug 3/4” MNPT Aluminum Zone 1&2   

B4951-1141

Plug M25 Aluminum Zone 1&2     

B4951-1142

Plug 3/4” MNPT CI D1&2  

The 

OXYvisor

 analyzer uses a quench fluorescence technique with a sensor optically isolated from the process, using the absorbency as a 

diagnostic function and analyzing the phase angle for measurement of the analyte, oxygen, in the modulated time domain.  This gives the analyzer the 
ability to measure accurately and precisely under various and changing ambient and process conditions.  

The resulting measurement is specific to oxygen concentration.  The luminophore is unaffected by other constituent gases and flow rate.  The 

measurement is applicable in both gas and liquid phase.  Temperature compensation is required to account for quenching efficiency at different 

temperatures and pressure compensation is required to measure at process pressured different than the pressure at time of calibration.

2.5 General principle of operation

OXYvisor

 Analyzer Overview

Figure 1 - 

OXYvisor

 overview of hardware features

2�6 

OXYvisor

 Analyzer Features  (Hardware/Firmware/Software)

OXYvisor

 Analyzer Dimensions

Figure 2 - 

OXYvisor

 Overall Dimensions

11.4"

288mm

DISPLAY AND THROUGH-
THE-GLASS TOUCH
KEYPAD

3/4" FNPT (X2)
or M25 x 1.5 6H 
conduit entries

1/2" FNPT (X1)

11.7"

298mm

FIBER OPTIC

BUSHING

P/N B3907-1SFP
or B3907-2SFP

JAM NUT
(FOR J-BOX ROTATION)

2 mm 
Hex Lock

ESC

OXYvisor

5.4"

136mm

10.5"

266mm

REAR WIRING
COMPARTMENT

2” MOUNT
BRACKET
B5008-1125

3.8"

97mm

ALTERNATE MOUNT STYLE

HORIZONTAL PIPE

ESC

OXYvisor

3/4" NPT (X4)
or M25 x 1.5 6H
conduit entries

FRONT VIEW

TOP VIEW

FRONT VIEW

(ALTERNATE MOUNT)

Rear (Wiring) Compartment 

Contains the connection 

terminals for power and all I/O’s 

except oxygen sensor.  The cap 

(cover) is threaded, sealed via 

O-ring and secured with a set 

screw.  The features allow easy, 

but controlled, access to wiring 

terminals.

Conduit Entries

There are four 3/4” FNPT or 

M25x1.5 6H conduit entries 

(size selected during ordering) 

for the 

Rear (Wiring) Compartment.

Conduit Entries (Junction Box)

There are two 3/4” FNPT or 

M25x1.5 6H conduit entries 

(size selected during ordering) 

for the 

Junction Box

Nameplate / Label 

Contains the part number, company 

info, and the 

Hazardous Area Markings.

Pipe Mount Bracket

316 SS, 2” Pipe / Field Mounting 

Bracket.  

Oxygen Sensor Entry

1/2” FNPT for the Oxygen sensor

Flameproof glass (Ex d)

Protects the HMI, LCD 

and other electronics

.

Human Machine Interface

The 

HMI

, located 

behind the flameproof 

glass, includes backlit 

LCD and 

Through-the- 

glass keypad

 for user 

configuration and full 

interactive use.

Junction Box 

Contains easily accesible sensor 

connection(s).

  

All analyzer models 

SMA connector for optical O2 sensor 

or optical fiber wand assembly.  

All analyzer models except Zone 1

Molex connector for Pt1000 RTD

Rugged Industrial Enclosure 

IP66 / NEMA4x, pressure die- 

casting with yellow chromating   

and chemically resistant paint.

Section 2

Technical Product Specification

Содержание BARBEN ANALYTICAL OXY visor

Страница 1: ...IOM Safety Installation Operation and Maintenance Manual Abridged Safety related part Optical Oxygen Analyzer BOA For use with BOSx Optical Sensors Optical O2 Products BOA ATEX IEC NEC CEC...

Страница 2: ...se of this analyzer However Barben Analytical assumes no responsibility for any inaccuracies that may be contained in this manual In no event will Barben Analytical be liable for direct indirect incid...

Страница 3: ...er 13 3 6 1 Protective Ground Earthing Screw 13 3 6 2 AC Powered Analyzer 13 3 6 3 DC Powered Analyzer 14 3 7 Wiring Input Outputs 14 3 8 Electrical Entries 15 3 9 Securing the Rear Wiring Compartment...

Страница 4: ...ne 1 HazLoc Installation Control DWG 2P0345 p 2 16 Figure 13 3 Zone 1 HazLoc Installation Control DWG 2P0345 p 3 17 Figure 14 1 Zone 2 HazLoc Installation Control DWG 2P0346 p 1 17 Figure 14 2 Zone 2...

Страница 5: ...r a factory qualified local representative see back page for complete contact information Section 1 General Information Safety 1 1 Introduction Purpose of this Manual 1 2 About the OXYvisor Product Ov...

Страница 6: ...al distributor located in Member State imports and takes ownership of the goods they will bear the responsibility for disposal of the goods according to the regulations of the own Member State Barben...

Страница 7: ...galvanic isolation 4 to 20 mA Active Linear or Bi Linear 24 Vdc Digital Outputs Alarm Relays 4 Programmable relays optically isolated passive 24Vdc 0 05A pilot duty 0 45 A general use resistive load...

Страница 8: ...r Overall Dimensions 11 4 288mm DISPLAY AND THROUGH THE GLASS TOUCH KEYPAD 3 4 FNPT X2 or M25 x 1 5 6H conduit entries 1 2 FNPT X1 11 7 298mm FIBER OPTIC BUSHING P N B3907 1SFP or B3907 2SFP JAM NUT F...

Страница 9: ...e 1 The product label identifies important information and attributes of the specific model and should be reviewed in detail prior to installation Class I Division 2 Groups A B C D T4A US NEC Standard...

Страница 10: ...594 0 689 RoHS Material 0 020 0 025 thick 316 Stainless Steel Hole Diameter 0 150 4 typ Corners R 0 125 4 typ Model Serial Power BOA DC 1 SFP B SI YYMXXXX 21 6 26 4 VDC 5 Watts 1 Product and Order Spe...

Страница 11: ...61010 1 July 31 st 2019 Carson City NV USA Date and Place of Issue Melvin W Thweatt Jr Product Line Manager 1906 AME BA OV DoC EU R1 Potentially Explosive Potentiellement Explosif Zone 2 Model Select...

Страница 12: ...Suggested tools to complete the installation are listed below M8 or Small crescent Wrench Used for 2 pipe mount bracket installation Large flathead Screwdriver Used to mount analyzer to mounting brack...

Страница 13: ...e analyzer enclosure must be connected Requirements for the Protective Ground conductor are as follows The protective conductor must be of equal or greater size than any other current carrying wiring...

Страница 14: ...ve PT100 PT1000 PT1000 Temperature Sensor PT1000 positive wire then jumper to PT100 terminal below PT100 V PT1000 4 wire 3 rd wire or jumper to above terminal PT100 4 wire RTD input Drive Voltage posi...

Страница 15: ...pose Class I Div 2 or Zone II 3G and BOSx SENSOR WITH INTEGRAL RTD ANALYZER RTD MOLEX CONNECTION IN J BOX BOSx OXYGEN SENSOR RTD MOLEX CONNECTION IN J BOX BOSx OXYGEN SENSOR WITH INTEGRAL RTD SEE BOSx...

Страница 16: ...NO FTZ 09 0031U IEC Ex MARKING Ex db IIC Gb x JUNCTION BOX MANUFACTURER IS LIMATHERM S A CONTROL DRAWING 2 Z L2453 COATING SPRAYED POLYURETHANE ON SPRAYED EPOXY LAYER THE BOTTOM JUNCTION BOX PORT IS...

Страница 17: ...Control DWG 2P0346 p 1 1 OF 3 C SHIELDS OXYvisor HAZARDOUS LOCATION INSTALLATION DRAWING ZONE 2 UNLESS OTHERWISE SPECIFIED ALL 015 FRACTIONAL 000 005 00 010 0 015 APVD BY DRAWN BY DRAWING NO DATE SCA...

Страница 18: ...LE SIZE SHEET PART NO MATERIAL DESCRIPTION APVD DWN DESCRIPTION DATE REV REVISIONS ANGLES 1 4 TOLERANCES DIMENSIONS IN INCHES ECO 5200 Convair Drive Carson City NV 89706 PH 775 883 2500 FAX 775 297 47...

Страница 19: ...4 M5 SPU CW OR XD DR120WIN 4 M25 M25 M25 M25 M24 M5 SPU CW ATEX CERTIFICATION NO FTZ 08 ATEX 0182U ATEX MARKING II 2G Ex db IIC Gb IEC Ex CERTIFICATION NO FTZ 09 0031U IEC Ex MARKING Ex db IIC Gb x JU...

Страница 20: ...S I DIVISION 2 HAZARDOUS AREA 1 Uo 2 5 V Vmax Uo Io 0 5 mA Imax Io Co 4 92 nF Ci Ccable Co Lo 16 40 H Li Lcable Lo USER SUPPLIED WIRING AND COMPONENTS 1 WIRING TO BE INSTALLED PER LOCAL CODE WIRING MU...

Страница 21: ...unauthorized operation was performed over Modbus Check the Modbus master device If a Modbus to USB cable is being used check cable requirements and replace N36 Unauthorized register was accessed via...

Страница 22: ...gone a reset undergone a reset Power OFF and ON If error is still present consult the factory Power OFF and ON If error is still present consult the factory X34 X34 This error indicates that some comp...

Страница 23: ...yzer Danger of explosion in areas subject to explosion hazard Connect the device correctly after maintenance Close and seal the device after maintenance work Flameproof joints are not intended to be r...

Страница 24: ...s nearly 14 000 colleagues at over 120 manufacturing locations around the world Supporting those operations are more than 80 sales and service locations across the United States and in more than 30 ot...

Отзывы: