IOM (Safety) - 

Installation, Operation and Maintenance

                      Optical Oxygen Analyzer

2.7 Hazardous Area Agency Certification & Markings - [Recapitulation]

2.7.1 Zone 1 Group IIC with Specific Conditions of Use

2.7.2 Zone 2 Group IIC with Specific Conditions of Use

OXYvisor

 is certified as 

Process Control Equipment

 for use in 

Zone Classified Hazardous Locations

 (

QVAJ

 and 

QVAJ7

) - Zone 1, Group IIC

Specific Conditions of Use 

- Consult Drawing No. 2P0345 for complete installation details.

• 

Unused entries shall be closed with suitable certified Ex db blanking elements with a minimum IP66 rating

• 

Flameproof joints are not intended to be repaired.  For repair or replacement of any part, contact Barben Analytical.

0539

II 2 G                 Ex db op is IIC T4 Gb

                         Ex db op is IIC T4 Gb

Ta = -20 C to +55 C

Ta = -20 C to +55 C

Ta = -20 C to +55 C
Ta = -20 C to +55 C

Class 1 Zone 1  AEx db op is IIC T4 Gb
 

                 Ex db op is IIC T4 Gb

Certificate Number:                                   

DEMKO 19 ATEX 2031,               issued by UL DEMKO International A/S

Certificate Number:             

  

IECEx UL 19.0040 issued by UL LLC

ATEX Zone 1, Group IIC Markings:

IECEx Zone 1, Group IIC Markings:

North American Zone 1, Group IIC Markings:

Standards: 

 

   EN 60079-0:2012+A11:2013

    EN 

60079-1:2014

    EN 

60079-28:2015

Standards:    

 

 

IEC 60079-0, 

6th  Edition,  Issued 2011-06

 

      

 

 

IEC 60079-1,    

7th  Edition,  Issued 2014-06 

 

      

 

 

IEC 60079-28,  

2nd Edition,  Issued 2015-05

US NEC Standards:  

 

 

UL   60079-0,  

 

6

th

  Edition,   Dated 2013-07-26

 

 

 

 

UL   60079-1,    

 

7

th

  Edition,   Dated 2015-09-18 

    UL 

 

 

60079-28, 

 

 2

nd

  Edition,  Dated 2017-09-15

CAN CEC Standards:  

 

 

CSA C22.2 No. 60079-0,   

3rd Edition,  Dated 2015-10

 

 

 

 

CSA C22.2 No. 60079-1,   

3rd Edition,  Dated 2016-05

 

 

 

 

CSA C22.2 No. 60079-28, 

1st  Edition,  Dated 2016-12

2.8 Product Label / Name Plate 

2.8.1 - Product Label / Name Plate Overview

2.7.3 Class I, Division 2 Groups A,B,C, & D 

The Product Label (Name Plate) is found at the top of the analyzer (Figure 1). The product label identifies important information and attributes of the 

specific model and should be reviewed in detail prior to installation. 

Class I, Division 2, Groups A, B, C, & D  T4A

US NEC Standards:  

 

                    UL 12.12.01 - Edition 9 Issued 2017/09/15

CAN CEC Standards: 

 

                    CSA C22.2 No. 213-17 - Edition 3 Issued 2017/09/15

The 

OXYvisor

 is certified as 

Process Control Equipment

 for use in hazardous locations (

QUZW

QUZW7

Class I, Division 2 Groups A, B, C, & D

!

WARNING

!

WARNING

Section 2

Technical Product Specification

The 

OXYvisor

 is certified as 

Process Control Equipment

 for use in 

Zone Classified Hazardous Locations 

- Zone 2, Group IIC

Specific Conditions of Use

 

- Consult Drawing No. 2P0346 for complete installation details.

• 

Transient protection shall be provided that is set at a level not exceeding 140% of the peak rated voltage value at the supply terminals to the equipment.

• 

The equipment shall only be used in an area of not more than Pollution Degree 2, as defined in IEC/EN 60664-1.

II 3 G                 Ex ec [ic] op is IIC T4 Gc

Ta = -20 C to +55 C

Certificate Number: 

 

 

DEMKO 19 ATEX 2036

ATEX Zone 2 Markings:

Standards:    

 

 

CENELEC    

EN 60079-0  +A11 

2012, 2013

    CENELEC 

 

 EN 

60079-7 

+A1 

2015, 

2018

    CENELEC 

 

 EN 

60079-11 

 2012

    CENELEC    

EN 60079-28 

2015

                           

 

       Ex ec [ic] op is IIC T4 Gc

Class 1 Zone 2 AEx ec [ic] op is IIC T4 Gc

Class 1 Zone 2   Ex ec [ic] op is IIC T4 Gc

Certificate Number: 

 

IECEx UL 19.0072

North American Zone 2, Group IIC Markings:

Standards:    

 

 

IEC 60079-0, 

 

Edition 6, 

Issued 2014-10

    IEC 

60079-7,  Edition 

5.1, 

Issued 

2014-06 

    IEC 

60079-11, 

 Edition 

6, 

Issued 

2014-10

    IEC 

60079-28, 

 Edition 

2, 

Issued 

2015-05

    IEC 

61010-2-201 

 Edition 

1, 

Issued 

2013

US NEC Standards:  

 

 

UL   60079-0,     

 

Edition 6, 

Dated 2017/10/20

 

 

 

 

UL   60079-7, 

 

Edition 5.1,  Dated 2017/04/21

 

 

 

 

UL   60079-11, 

 

Edition 6, 

Dated 2014/03/28

 

 

 

 

UL   60079-28, 

 

Edition 2, 

Dated 2017/12/08

CAN CEC Standards: 

   

 

CSA C22.2 No. 60079-0,  Edition 3,  

Issue Date 2015/10/01

 

 

 

 

CSA C22.2 No. 60079-7,   Edition 2,  

Issue Date 2016

 

 

 

 

CSA C22.2 No. 60079-11,  Edition 2,  

Issue Date 2014/02/01

 

 

 

 

CSA C22.2 No. 60079-28, Edition 1,  

Issue Date 2016

Содержание BARBEN ANALYTICAL OXY visor

Страница 1: ...IOM Safety Installation Operation and Maintenance Manual Abridged Safety related part Optical Oxygen Analyzer BOA For use with BOSx Optical Sensors Optical O2 Products BOA ATEX IEC NEC CEC...

Страница 2: ...se of this analyzer However Barben Analytical assumes no responsibility for any inaccuracies that may be contained in this manual In no event will Barben Analytical be liable for direct indirect incid...

Страница 3: ...er 13 3 6 1 Protective Ground Earthing Screw 13 3 6 2 AC Powered Analyzer 13 3 6 3 DC Powered Analyzer 14 3 7 Wiring Input Outputs 14 3 8 Electrical Entries 15 3 9 Securing the Rear Wiring Compartment...

Страница 4: ...ne 1 HazLoc Installation Control DWG 2P0345 p 2 16 Figure 13 3 Zone 1 HazLoc Installation Control DWG 2P0345 p 3 17 Figure 14 1 Zone 2 HazLoc Installation Control DWG 2P0346 p 1 17 Figure 14 2 Zone 2...

Страница 5: ...r a factory qualified local representative see back page for complete contact information Section 1 General Information Safety 1 1 Introduction Purpose of this Manual 1 2 About the OXYvisor Product Ov...

Страница 6: ...al distributor located in Member State imports and takes ownership of the goods they will bear the responsibility for disposal of the goods according to the regulations of the own Member State Barben...

Страница 7: ...galvanic isolation 4 to 20 mA Active Linear or Bi Linear 24 Vdc Digital Outputs Alarm Relays 4 Programmable relays optically isolated passive 24Vdc 0 05A pilot duty 0 45 A general use resistive load...

Страница 8: ...r Overall Dimensions 11 4 288mm DISPLAY AND THROUGH THE GLASS TOUCH KEYPAD 3 4 FNPT X2 or M25 x 1 5 6H conduit entries 1 2 FNPT X1 11 7 298mm FIBER OPTIC BUSHING P N B3907 1SFP or B3907 2SFP JAM NUT F...

Страница 9: ...e 1 The product label identifies important information and attributes of the specific model and should be reviewed in detail prior to installation Class I Division 2 Groups A B C D T4A US NEC Standard...

Страница 10: ...594 0 689 RoHS Material 0 020 0 025 thick 316 Stainless Steel Hole Diameter 0 150 4 typ Corners R 0 125 4 typ Model Serial Power BOA DC 1 SFP B SI YYMXXXX 21 6 26 4 VDC 5 Watts 1 Product and Order Spe...

Страница 11: ...61010 1 July 31 st 2019 Carson City NV USA Date and Place of Issue Melvin W Thweatt Jr Product Line Manager 1906 AME BA OV DoC EU R1 Potentially Explosive Potentiellement Explosif Zone 2 Model Select...

Страница 12: ...Suggested tools to complete the installation are listed below M8 or Small crescent Wrench Used for 2 pipe mount bracket installation Large flathead Screwdriver Used to mount analyzer to mounting brack...

Страница 13: ...e analyzer enclosure must be connected Requirements for the Protective Ground conductor are as follows The protective conductor must be of equal or greater size than any other current carrying wiring...

Страница 14: ...ve PT100 PT1000 PT1000 Temperature Sensor PT1000 positive wire then jumper to PT100 terminal below PT100 V PT1000 4 wire 3 rd wire or jumper to above terminal PT100 4 wire RTD input Drive Voltage posi...

Страница 15: ...pose Class I Div 2 or Zone II 3G and BOSx SENSOR WITH INTEGRAL RTD ANALYZER RTD MOLEX CONNECTION IN J BOX BOSx OXYGEN SENSOR RTD MOLEX CONNECTION IN J BOX BOSx OXYGEN SENSOR WITH INTEGRAL RTD SEE BOSx...

Страница 16: ...NO FTZ 09 0031U IEC Ex MARKING Ex db IIC Gb x JUNCTION BOX MANUFACTURER IS LIMATHERM S A CONTROL DRAWING 2 Z L2453 COATING SPRAYED POLYURETHANE ON SPRAYED EPOXY LAYER THE BOTTOM JUNCTION BOX PORT IS...

Страница 17: ...Control DWG 2P0346 p 1 1 OF 3 C SHIELDS OXYvisor HAZARDOUS LOCATION INSTALLATION DRAWING ZONE 2 UNLESS OTHERWISE SPECIFIED ALL 015 FRACTIONAL 000 005 00 010 0 015 APVD BY DRAWN BY DRAWING NO DATE SCA...

Страница 18: ...LE SIZE SHEET PART NO MATERIAL DESCRIPTION APVD DWN DESCRIPTION DATE REV REVISIONS ANGLES 1 4 TOLERANCES DIMENSIONS IN INCHES ECO 5200 Convair Drive Carson City NV 89706 PH 775 883 2500 FAX 775 297 47...

Страница 19: ...4 M5 SPU CW OR XD DR120WIN 4 M25 M25 M25 M25 M24 M5 SPU CW ATEX CERTIFICATION NO FTZ 08 ATEX 0182U ATEX MARKING II 2G Ex db IIC Gb IEC Ex CERTIFICATION NO FTZ 09 0031U IEC Ex MARKING Ex db IIC Gb x JU...

Страница 20: ...S I DIVISION 2 HAZARDOUS AREA 1 Uo 2 5 V Vmax Uo Io 0 5 mA Imax Io Co 4 92 nF Ci Ccable Co Lo 16 40 H Li Lcable Lo USER SUPPLIED WIRING AND COMPONENTS 1 WIRING TO BE INSTALLED PER LOCAL CODE WIRING MU...

Страница 21: ...unauthorized operation was performed over Modbus Check the Modbus master device If a Modbus to USB cable is being used check cable requirements and replace N36 Unauthorized register was accessed via...

Страница 22: ...gone a reset undergone a reset Power OFF and ON If error is still present consult the factory Power OFF and ON If error is still present consult the factory X34 X34 This error indicates that some comp...

Страница 23: ...yzer Danger of explosion in areas subject to explosion hazard Connect the device correctly after maintenance Close and seal the device after maintenance work Flameproof joints are not intended to be r...

Страница 24: ...s nearly 14 000 colleagues at over 120 manufacturing locations around the world Supporting those operations are more than 80 sales and service locations across the United States and in more than 30 ot...

Отзывы: