IOM (Safety) - 

Installation, Operation and Maintenance

                      Optical Oxygen Analyzer

Unused entries shall be closed with suitable certified Ex db blanking elements with a minimum IP66 rating

After wiring the power and any input / output connections and before powering up the analyzer, the Rear (Wiring) Compartment lid should 
be installed and fully tightened.  Follow the detailed instruction found in Section 5.4.1.

3�8 Electrical Entries

3.9 Securing the Rear (Wiring) Compartment lid.

Section 3

Installation

The mating RTD Molex connection in the Junction Box is only available on 

OXYvisor

 models with Agency Approval selection “2” 

(cULus CID2 and ATEX/IEC Zone 2 3G) per 

OXYvisor

 Configuration - Product Model Selection, Table 2.  

Barben Optical Sensors (BOS) connect to the 

OXYvisor

 through a fiber optic cable connection.  The fiber optic cable from the 

OXYvisor

 is easily 

mated to the BOS oxygen sensor via a finger-tightened SMA coupler connection, located in the lower sensor junction box (J-Box).  Both integral and 

external RTD’s can be connected to the 

OXYvisor

 through the Rear (Wiring) Compartment and via the J-Box when the later connection is available.  

Figures 10, 11 & 12 show some common arrangements for the 

OXYvisor

 and BOS sensors.

3.10.1 BOS Sensor with Integral and External RTD via J-Box

3.10 Sensor Connection Options

Figure 10 - BOS Sensor With Integral RTD Through Junction Box

When the BOS sensor is specified with an integral RTD, and the 

OXYvisor

 model is selected with Agency Approval “2” or “3”, the RTD can also be 

connected through the J-Box via a Molex connector. See Figure 10.

ESC

OXYvisor

BOSx-FLEXSENSE

Oxygen

01/06/2017 10:55 AM

850.9 mBar
67.9°F
Err: 0

20.946

„‡‡‡

% O2

EXAMPLE - GENERAL ARRANGEMENT 

OF OXYVISOR (General Purpose / Class I 

Div. 2 / or Zone II 3G) and BOSx SENSOR 

WITH INTEGRAL RTD

ANALYZER RTD 

MOLEX CONNECTION 

IN J-BOX 

BOSx OXYGEN SENSOR

RTD MOLEX CONNECTION 

IN J-BOX 

BOSx OXYGEN SENSOR 

WITH INTEGRAL RTD

(SEE BOSx DATA 

SHEET FOR OPTIONS)

ANALYZER FIBER

OPTIC SMA 

CONNECTION IN 

J-BOX 

FIBER OPTIC SMA 

TO SMA COUPLER

P/N: B3907-1122 

BOSx OXYGEN 

SENSOR FIBER 

OPTIC SMA 

CONNECTION IN

J-BOX 

Figure 11 - BOS Sensor And External RTD through Junction Box

ESC

OXYvisor

BOSx-FLEXSENSE

PT1000 RTD

Oxygen

01/06/2017 10:55 AM

850.9 mBar
67.9°F
Err: 0

20.946

„‡‡‡

% O2

EXAMPLE - GENERAL ARRANGEMENT 

OF OXYVISOR (General Purpose / Class I 

Div. 2 / or Zone II 3G) and BOSx SENSOR 

WITH EXTERNAL RTD

ANALYZER RTD 

MOLEX CONNECTION 

IN J-BOX 

BOSx OXYGEN SENSOR

RTD MOLEX CONNECTION 

IN J-BOX 

ANALYZER FIBER

OPTIC SMA 

CONNECTION IN 

J-BOX 

FIBER OPTIC SMA 

TO SMA COUPLER

P/N: B3907-1122 

BOSx OXYGEN 

SENSOR FIBER 

OPTIC SMA 

CONNECTION IN

J-BOX 

BOSx OXYGEN SENSOR 

WITHOUT RTD

(SEE BOSx DATA 

SHEET FOR OPTIONS)

EXTERNAL RTD

3.10.3 BOS Sensor with External RTD via Rear (Wiring) Compartment

Figure 12 - BOS Sensor And External RTD or Zone 1

For 

OXYvisor

 installations in Zone 1, where the model is selected with Agency Approval “1”, the BOS sensor (as standard) is not allowed to be 

installed with integral RTD.  An external RTD will be the standard recommended installation.  For this case, the RTD can be wired through the Rear 
(Wiring) Compartment as shown in Figure 12. 

ESC

OXYvisor

BOSx-FLEXSENSE

RIGID CONDUIT or Ex d GLANDS & CABLING

BOSx OXYGEN SENSOR 

(SEE BOSx DATASHEET

FOR OPTIONS)

 

USER SUPPLIED RTD OR 

TEMPERATURE TRANSMITTER

CERTIFIED FOR HAZARDOUS 

AREA

EXAMPLE - GENERAL ARRANGEMENT 

OF OXYVISOR (Zone I 2G) and BOSx SENSOR 

WITH EXTERNAL RTD

ANALYZER FIBER

OPTIC SMA 

CONNECTION IN 

J-BOX 

FIBER OPTIC SMA 

TO SMA COUPLER

P/N: B3907-1122 

BOSx OXYGEN 

SENSOR FIBER 

OPTIC SMA 

CONNECTION IN

J-BOX 

Oxygen

01/06/2017 10:55 AM

850.9 mBar
67.9°F
Err: 0

20.946

„‡‡‡

% O2

!

CAUTION

!

CAUTION

USEFUL INFORMATION

Содержание BARBEN ANALYTICAL OXY visor

Страница 1: ...IOM Safety Installation Operation and Maintenance Manual Abridged Safety related part Optical Oxygen Analyzer BOA For use with BOSx Optical Sensors Optical O2 Products BOA ATEX IEC NEC CEC...

Страница 2: ...se of this analyzer However Barben Analytical assumes no responsibility for any inaccuracies that may be contained in this manual In no event will Barben Analytical be liable for direct indirect incid...

Страница 3: ...er 13 3 6 1 Protective Ground Earthing Screw 13 3 6 2 AC Powered Analyzer 13 3 6 3 DC Powered Analyzer 14 3 7 Wiring Input Outputs 14 3 8 Electrical Entries 15 3 9 Securing the Rear Wiring Compartment...

Страница 4: ...ne 1 HazLoc Installation Control DWG 2P0345 p 2 16 Figure 13 3 Zone 1 HazLoc Installation Control DWG 2P0345 p 3 17 Figure 14 1 Zone 2 HazLoc Installation Control DWG 2P0346 p 1 17 Figure 14 2 Zone 2...

Страница 5: ...r a factory qualified local representative see back page for complete contact information Section 1 General Information Safety 1 1 Introduction Purpose of this Manual 1 2 About the OXYvisor Product Ov...

Страница 6: ...al distributor located in Member State imports and takes ownership of the goods they will bear the responsibility for disposal of the goods according to the regulations of the own Member State Barben...

Страница 7: ...galvanic isolation 4 to 20 mA Active Linear or Bi Linear 24 Vdc Digital Outputs Alarm Relays 4 Programmable relays optically isolated passive 24Vdc 0 05A pilot duty 0 45 A general use resistive load...

Страница 8: ...r Overall Dimensions 11 4 288mm DISPLAY AND THROUGH THE GLASS TOUCH KEYPAD 3 4 FNPT X2 or M25 x 1 5 6H conduit entries 1 2 FNPT X1 11 7 298mm FIBER OPTIC BUSHING P N B3907 1SFP or B3907 2SFP JAM NUT F...

Страница 9: ...e 1 The product label identifies important information and attributes of the specific model and should be reviewed in detail prior to installation Class I Division 2 Groups A B C D T4A US NEC Standard...

Страница 10: ...594 0 689 RoHS Material 0 020 0 025 thick 316 Stainless Steel Hole Diameter 0 150 4 typ Corners R 0 125 4 typ Model Serial Power BOA DC 1 SFP B SI YYMXXXX 21 6 26 4 VDC 5 Watts 1 Product and Order Spe...

Страница 11: ...61010 1 July 31 st 2019 Carson City NV USA Date and Place of Issue Melvin W Thweatt Jr Product Line Manager 1906 AME BA OV DoC EU R1 Potentially Explosive Potentiellement Explosif Zone 2 Model Select...

Страница 12: ...Suggested tools to complete the installation are listed below M8 or Small crescent Wrench Used for 2 pipe mount bracket installation Large flathead Screwdriver Used to mount analyzer to mounting brack...

Страница 13: ...e analyzer enclosure must be connected Requirements for the Protective Ground conductor are as follows The protective conductor must be of equal or greater size than any other current carrying wiring...

Страница 14: ...ve PT100 PT1000 PT1000 Temperature Sensor PT1000 positive wire then jumper to PT100 terminal below PT100 V PT1000 4 wire 3 rd wire or jumper to above terminal PT100 4 wire RTD input Drive Voltage posi...

Страница 15: ...pose Class I Div 2 or Zone II 3G and BOSx SENSOR WITH INTEGRAL RTD ANALYZER RTD MOLEX CONNECTION IN J BOX BOSx OXYGEN SENSOR RTD MOLEX CONNECTION IN J BOX BOSx OXYGEN SENSOR WITH INTEGRAL RTD SEE BOSx...

Страница 16: ...NO FTZ 09 0031U IEC Ex MARKING Ex db IIC Gb x JUNCTION BOX MANUFACTURER IS LIMATHERM S A CONTROL DRAWING 2 Z L2453 COATING SPRAYED POLYURETHANE ON SPRAYED EPOXY LAYER THE BOTTOM JUNCTION BOX PORT IS...

Страница 17: ...Control DWG 2P0346 p 1 1 OF 3 C SHIELDS OXYvisor HAZARDOUS LOCATION INSTALLATION DRAWING ZONE 2 UNLESS OTHERWISE SPECIFIED ALL 015 FRACTIONAL 000 005 00 010 0 015 APVD BY DRAWN BY DRAWING NO DATE SCA...

Страница 18: ...LE SIZE SHEET PART NO MATERIAL DESCRIPTION APVD DWN DESCRIPTION DATE REV REVISIONS ANGLES 1 4 TOLERANCES DIMENSIONS IN INCHES ECO 5200 Convair Drive Carson City NV 89706 PH 775 883 2500 FAX 775 297 47...

Страница 19: ...4 M5 SPU CW OR XD DR120WIN 4 M25 M25 M25 M25 M24 M5 SPU CW ATEX CERTIFICATION NO FTZ 08 ATEX 0182U ATEX MARKING II 2G Ex db IIC Gb IEC Ex CERTIFICATION NO FTZ 09 0031U IEC Ex MARKING Ex db IIC Gb x JU...

Страница 20: ...S I DIVISION 2 HAZARDOUS AREA 1 Uo 2 5 V Vmax Uo Io 0 5 mA Imax Io Co 4 92 nF Ci Ccable Co Lo 16 40 H Li Lcable Lo USER SUPPLIED WIRING AND COMPONENTS 1 WIRING TO BE INSTALLED PER LOCAL CODE WIRING MU...

Страница 21: ...unauthorized operation was performed over Modbus Check the Modbus master device If a Modbus to USB cable is being used check cable requirements and replace N36 Unauthorized register was accessed via...

Страница 22: ...gone a reset undergone a reset Power OFF and ON If error is still present consult the factory Power OFF and ON If error is still present consult the factory X34 X34 This error indicates that some comp...

Страница 23: ...yzer Danger of explosion in areas subject to explosion hazard Connect the device correctly after maintenance Close and seal the device after maintenance work Flameproof joints are not intended to be r...

Страница 24: ...s nearly 14 000 colleagues at over 120 manufacturing locations around the world Supporting those operations are more than 80 sales and service locations across the United States and in more than 30 ot...

Отзывы: