background image

10

IOM (Safety) - 

Installation, Operation and Maintenance

Optical Oxygen Analyzer

Figure 3 - 

OXYvisor

 Product Label / Name Plate

 OXYvisor

II 2 G      Ex db op is IIC T4 Gb

    Ex db op is IIC T4 Gb

Ta = -20°C to +55°C, IP 66, NEMA 4x

 Class I  Zone 1   AEx db op is IIC T4 Gb

     Ex db op is IIC T4 Gb

Installation DWG# 2P0345

0539

IECEx UL 18.0033

DEMKO 18 ATEX 2031

AVERTISSEMENT:

 

Ne pas ouvrir quand une atmosphère 

explosive est présente. Sceller toutes les sorties de câbles 

conformément aux codes gouvernementaux et locaux.

WARNING: Do not open when an explosive 

atmosphere is present.  Seal all cable exits in 

accordance with government and local code.

!

!

Made in Carson City NV - USA

www.BarbenAnalytical.com

Model#
Serial #
Power: 

RoHS

!

WARNING: Consult Conditions of Safe Use in installation 

manual and DWG# 2P0345 to preserve the protection 

afforded by the equipment.

!

AVERTISSEMENT: Consultez les conditions d'utilisation 

sécurisée du manuel d'installation et le DWG n° 2P0345 

pour préserver la protection offerte par l'équipement.

BOA-DC-1-SFP-B-SI
YYMXXXX
21.6 - 26.4 VDC

               5 Watts

E502700 Process Control Equipment for 

Use in Zone Classified Hazardous Locations

Branding Logo

Parent Company Logo

 Product Name

Serial # contains the 

date of manufacture.  

The first three numbers 

are YYM (YY=Year, M= 

month by alphabet, 

less”i”) then XXXX = 

sequential build)

Power: Acceptable 

range, power type and 

max wattage drawn

Installation Drawing contains 

critical information about the 

installation requirements.  This 

must be reviewed prior to 

installation.

Important Warnings! that 

should be adhered to at all 

times.

Hazardous Area Markings

2.8.2 - Zone 1 Groups IIC - Barben Analytical | AMETEK

2.8.3 - Zone 2 Groups IIC - Barben Analytical | AMETEK

2V0138

05/24/2018

10/31/2018

Z1 Final submittal

1 OF 1

1 : 1

INSTRUMENT

M.Thweatt

B. Wolf

A

B5001-1097

SS Nameplate Z1 (IEC/ATEX/UL/cUL)

OXYVISOR - BOA

  B    

UNLESS

OTHERWISE

SPECIFIED

FOR

PART NO.

APVD BY

DRAWN BY

DRAWING NO

DATE

SCALE

SIZE

SHEET

APVD

DWN

MT

DESCRIPTION

DATE

REV

REVISIONS

DIMENSIONS

IN mm

ECO

5200 Convair Drive Carson City, NV 89706 PH(775)883-2500 FAX (775)297-4740

BW

07/19/2019 Z1 Final - UL updates - CE Move

  C

MT

BW

11/07/2021 S/N Schema Change & Ex Sched 

  D

MT

BW

 OXYvisor

     II 2 G Ex db op is IIC T4 Gb

Ta = -20°C to +55°C, IP 66, NEMA 4x

 Class I  Zone 1   AEx db op is IIC T4 Gb

     Ex db op is IIC T4 Gb

0539

IECEx UL 19.0040

DEMKO 19 ATEX 2031

AVERTISSEMENT:

 Ne pas ouvrir quand une atmosphère 

explosive est présente. Sceller toutes les sorties de câbles 

conformément aux codes gouvernementaux et locaux.

WARNING: Do not open when an explosive 

atmosphere is present.  Seal all cable exits in 

accordance with government and local code.

!

!

5200 Convair Drive, Carson City NV 89706 - USA

www.BarbenAnalytical.com

Model#

Serial #

Power: 

!

WARNING: Consult Conditions of Safe Use in installation 

manual and DWG# 2P0345 to preserve the protection 

afforded by the equipment.

!

AVERTISSEMENT: Consultez les conditions d'utilisation 

sécurisée du manuel d'installation et le DWG n° 2P0345 

pour préserver la protection offerte par l'équipement.

Process Control Equipment for Use in 

Zone Classified Hazardous Locations

E502700

1

4.40”

3.15”

3.539”

2.851”

0.563”

0.865”

1.594”

0.689”

RoHS

Material 

= 0.020 - 0.025” thick 

316 Stainless Steel

Hole Diameter

 =   0.150” (4-typ)

Corners 

R =  

   0.125” (4-typ)

Model#
Serial #
Power: 

BOA-DC-1-SFP-B-SI
YYMXXXX
21.6 - 26.4 VDC

               5 Watts

1

Product and Order Specifc Information (Model#, Serial# and 

Power) to be engraved at Barben Facility at time of order

Model# 

- Approved Product Nomenclature 

OXYvisor Optical Oxygen Analyzer, Series BOA; followed by -AC or -DC; followed by -1; 

followed by -S025, -S050, -S100- Sxyz*, or -SFP; followed by -B, -L, or -R; followed by -SI or 

- AM (*Where xyz = fiber length in tenth of meters)

Serial #

   - YYMXXXX (YY = Year, M = Month by alphabet, less “i”,  XXXX = sequential build)

Power 

  -  (a) Nature of supply

 

 

1)  a.c.:  symbol 2, Rated Mains frequency or range of frequency

 

 

2.) d.c:   symbol 1 of Table 1

 

(b) Rated range of supply voltage 

 

(c) Maximum Rated Power in Watts or VA

Example of Product and Order Specific Info engraved by manufacturer:

- NOTICE - 

EX SCHEDULE DRAWING

NO REVISION WITHOUT AGENCY AND 

EX PERSONNEL AUTHORIZATION

Section 2

Technical Product Specification

2V0148

06/26/2019

1 OF 1

1 : 1

INSTRUMENT

M.Thweatt

B. Wolf

A

B5001-1112

SS Nameplate Z2 (IEC/ATEX/NEC)

OXYvisor - BOA

UNLESS

OTHERWISE

SPECIFIED

FOR

PART NO.

APVD BY

DRAWN BY

DRAWING NO

DATE

SCALE

SIZE

SHEET

APVD

DWN

DESCRIPTION

DATE

REV

REVISIONS

DIMENSIONS

IN mm

ECO

5200 Convair Drive Carson City, NV 89706 PH(775)883-2500 FAX (775)297-4740

 OXYvisor

 II 3 G  Ex ec [ic] op is IIC T4 Gc

Ta = -20°C to +55°C, IP 66, NEMA 4x

Class I  Zone 2

  AEx ec [ic] op is IIC T4 Gc

     Ex ec [ic] op is IIC T4 Gc

IECEx UL 19.0072

DEMKO 19 ATEX 2036

AVERTISSEMENT:

 Ne pas ouvrir quand une atmosphère 

explosive est présente. Sceller toutes les sorties de câbles 

conformément aux codes gouvernementaux et locaux.

WARNING: Do not open when an explosive 

atmosphere is present.  Seal all cable exits in 

accordance with government and local code.

!

!

5200 Convair Drive, Carson City NV 89706 - USA

www.BarbenAnalytical.com

Model#

Serial #

Power: 

!

WARNING: Consult Conditions of Safe Use in installation 

manual and DWG# 2P0346 to preserve the protection 

afforded by the equipment.

!

AVERTISSEMENT: Consultez les conditions d'utilisation 

sécurisée du manuel d'installation et le DWG n° 2P0346 

pour préserver la protection offerte par l'équipement.

Process Control Equipment for Use in 

Zone Classified Hazardous Locations

E479864

1

RoHS

4.40”

3.15”

3.539”

2.851”

0.563”

0.865”

1.594”

0.689”

Material 

= 0.020 - 0.025” thick 

316 Stainless Steel

Hole Diameter

 =   0.150” (4-typ)

Corners 

R =  

   0.125” (4-typ)

Model#
Serial #
Power: 

BOA-DC-3-SFP-B-SI
YYMXXXX
21.6 - 26.4 VDC

               5 Watts

1

Product and Order Specifc Information (Model#, Serial# and 

Power) to be engraved at Barben Facility at time of order

Model# 

- Approved Product Nomenclature 

OXYvisor Optical Oxygen Analyzer, Series BOA; followed by -AC or -DC; followed by -3; 

followed by -S025, -S050, -S100- Sxyz*, or -SFP; followed by -B, -L, or -R; followed by -SI or 

- AM (*Where xyz = fiber length in tenth of meters)

Serial #

   - YYMXXXX (YY = Year, M = Month by alphabet, less “i”,  XXXX = sequential build)

Power 

  -  (a) Nature of supply

 

 

1)  a.c.:  symbol 2, Rated Mains frequency or range of frequency

 

 

2.) d.c:   symbol 1 of Table 1

 

(b) Rated range of supply voltage 

 

(c) Maximum Rated Power in Watts or VA

Example of Product and Order Specific Info engraved by manufacturer:

- NOTICE - 

EX SCHEDULE DRAWING

NO REVISION WITHOUT AGENCY AND 

EX PERSONNEL AUTHORIZATION

06/26/2019 Zone 2 - BOA - Release 

  A

MT

BW

11/07/2021 S/N Schema Change & Ex Sched 

  B

MT

BW

Содержание BARBEN ANALYTICAL OXY visor

Страница 1: ...IOM Safety Installation Operation and Maintenance Manual Abridged Safety related part Optical Oxygen Analyzer BOA For use with BOSx Optical Sensors Optical O2 Products BOA ATEX IEC NEC CEC...

Страница 2: ...se of this analyzer However Barben Analytical assumes no responsibility for any inaccuracies that may be contained in this manual In no event will Barben Analytical be liable for direct indirect incid...

Страница 3: ...er 13 3 6 1 Protective Ground Earthing Screw 13 3 6 2 AC Powered Analyzer 13 3 6 3 DC Powered Analyzer 14 3 7 Wiring Input Outputs 14 3 8 Electrical Entries 15 3 9 Securing the Rear Wiring Compartment...

Страница 4: ...ne 1 HazLoc Installation Control DWG 2P0345 p 2 16 Figure 13 3 Zone 1 HazLoc Installation Control DWG 2P0345 p 3 17 Figure 14 1 Zone 2 HazLoc Installation Control DWG 2P0346 p 1 17 Figure 14 2 Zone 2...

Страница 5: ...r a factory qualified local representative see back page for complete contact information Section 1 General Information Safety 1 1 Introduction Purpose of this Manual 1 2 About the OXYvisor Product Ov...

Страница 6: ...al distributor located in Member State imports and takes ownership of the goods they will bear the responsibility for disposal of the goods according to the regulations of the own Member State Barben...

Страница 7: ...galvanic isolation 4 to 20 mA Active Linear or Bi Linear 24 Vdc Digital Outputs Alarm Relays 4 Programmable relays optically isolated passive 24Vdc 0 05A pilot duty 0 45 A general use resistive load...

Страница 8: ...r Overall Dimensions 11 4 288mm DISPLAY AND THROUGH THE GLASS TOUCH KEYPAD 3 4 FNPT X2 or M25 x 1 5 6H conduit entries 1 2 FNPT X1 11 7 298mm FIBER OPTIC BUSHING P N B3907 1SFP or B3907 2SFP JAM NUT F...

Страница 9: ...e 1 The product label identifies important information and attributes of the specific model and should be reviewed in detail prior to installation Class I Division 2 Groups A B C D T4A US NEC Standard...

Страница 10: ...594 0 689 RoHS Material 0 020 0 025 thick 316 Stainless Steel Hole Diameter 0 150 4 typ Corners R 0 125 4 typ Model Serial Power BOA DC 1 SFP B SI YYMXXXX 21 6 26 4 VDC 5 Watts 1 Product and Order Spe...

Страница 11: ...61010 1 July 31 st 2019 Carson City NV USA Date and Place of Issue Melvin W Thweatt Jr Product Line Manager 1906 AME BA OV DoC EU R1 Potentially Explosive Potentiellement Explosif Zone 2 Model Select...

Страница 12: ...Suggested tools to complete the installation are listed below M8 or Small crescent Wrench Used for 2 pipe mount bracket installation Large flathead Screwdriver Used to mount analyzer to mounting brack...

Страница 13: ...e analyzer enclosure must be connected Requirements for the Protective Ground conductor are as follows The protective conductor must be of equal or greater size than any other current carrying wiring...

Страница 14: ...ve PT100 PT1000 PT1000 Temperature Sensor PT1000 positive wire then jumper to PT100 terminal below PT100 V PT1000 4 wire 3 rd wire or jumper to above terminal PT100 4 wire RTD input Drive Voltage posi...

Страница 15: ...pose Class I Div 2 or Zone II 3G and BOSx SENSOR WITH INTEGRAL RTD ANALYZER RTD MOLEX CONNECTION IN J BOX BOSx OXYGEN SENSOR RTD MOLEX CONNECTION IN J BOX BOSx OXYGEN SENSOR WITH INTEGRAL RTD SEE BOSx...

Страница 16: ...NO FTZ 09 0031U IEC Ex MARKING Ex db IIC Gb x JUNCTION BOX MANUFACTURER IS LIMATHERM S A CONTROL DRAWING 2 Z L2453 COATING SPRAYED POLYURETHANE ON SPRAYED EPOXY LAYER THE BOTTOM JUNCTION BOX PORT IS...

Страница 17: ...Control DWG 2P0346 p 1 1 OF 3 C SHIELDS OXYvisor HAZARDOUS LOCATION INSTALLATION DRAWING ZONE 2 UNLESS OTHERWISE SPECIFIED ALL 015 FRACTIONAL 000 005 00 010 0 015 APVD BY DRAWN BY DRAWING NO DATE SCA...

Страница 18: ...LE SIZE SHEET PART NO MATERIAL DESCRIPTION APVD DWN DESCRIPTION DATE REV REVISIONS ANGLES 1 4 TOLERANCES DIMENSIONS IN INCHES ECO 5200 Convair Drive Carson City NV 89706 PH 775 883 2500 FAX 775 297 47...

Страница 19: ...4 M5 SPU CW OR XD DR120WIN 4 M25 M25 M25 M25 M24 M5 SPU CW ATEX CERTIFICATION NO FTZ 08 ATEX 0182U ATEX MARKING II 2G Ex db IIC Gb IEC Ex CERTIFICATION NO FTZ 09 0031U IEC Ex MARKING Ex db IIC Gb x JU...

Страница 20: ...S I DIVISION 2 HAZARDOUS AREA 1 Uo 2 5 V Vmax Uo Io 0 5 mA Imax Io Co 4 92 nF Ci Ccable Co Lo 16 40 H Li Lcable Lo USER SUPPLIED WIRING AND COMPONENTS 1 WIRING TO BE INSTALLED PER LOCAL CODE WIRING MU...

Страница 21: ...unauthorized operation was performed over Modbus Check the Modbus master device If a Modbus to USB cable is being used check cable requirements and replace N36 Unauthorized register was accessed via...

Страница 22: ...gone a reset undergone a reset Power OFF and ON If error is still present consult the factory Power OFF and ON If error is still present consult the factory X34 X34 This error indicates that some comp...

Страница 23: ...yzer Danger of explosion in areas subject to explosion hazard Connect the device correctly after maintenance Close and seal the device after maintenance work Flameproof joints are not intended to be r...

Страница 24: ...s nearly 14 000 colleagues at over 120 manufacturing locations around the world Supporting those operations are more than 80 sales and service locations across the United States and in more than 30 ot...

Отзывы: