IOM (Safety) -
Installation, Operation and Maintenance
Optical Oxygen Analyzer
Section 3
Installation
Figure 15.1 - Class I Div. 2 HazLoc - Installation Control DWG - 2P0335 (p.1)
3.11.3 Class I, Div2 - 2P0335
1 OF 3
C. SHIELDS
OXYvisor HAZARDOUS LOCATION
INSTALLATION DRAWING DIVISION 2
UNLESS
OTHERWISE
SPECIFIED
ALL ± .015"
FRACTIONAL
.000 ± .005"
.00 ± .010"
.0 ± .015"
APVD BY
DRAWN BY
DRAWING NO
DATE
SCALE
SIZE
SHEET
PART NO.
MATERIAL/DESCRIPTION
APVD
DWN
DESCRIPTION
DATE
REV
REVISIONS
ANGLES ± 1/4°
TOLERANCES
DIMENSIONS
IN INCHES
ECO#
5200 Convair Drive Carson City, NV 89706 PH(775)883-2500 FAX (775)297-4740
B. WOLF
01/26/2018
2P0335
-
OXYvisor
PART NO.
D
NTS
FOR
A 09/26/18 U.L. CHANGES CS BW
2
ALL USED CONDUIT PORTS IN ENCLOSURE AND JUNCTION BOX MUST BE SEALED WITH
APPROPRIATELY CERTIFIED FITTINGS AND SEALANTS PER THE APPLICABLE HAZARDOUS LOCATION.
ALL UNUSED PORTS MUST BE PLUGGED. TO ACHIEVE IP66 RATING (MAXIMUM,)
ALL CUSTOMER SUPPLIED FITTINGS, PLUGS, AND SEALANTS MUST BE SUITABLY RATED.
THREADED CONDUIT PORT OPTIONS:
M25 X 1.5 6H
3/4" NPT FEMALE
FRONT AND REAR ENCLOSURE DOORS MUST BE SECURED WITH THE PROVIDED
LOCKING SCREW TO PREVENT OPENING WITHOUT THE USE OF A TOOL.
(M4 LOCKING SCREW REQUIRES M2 HEX WRENCH)
ENCLOSURE MANUFACTURER IS LIMATHERM S.A.:
CONTROL DRAWING: 1-Z-L4249
COATING: SPRAYED POLYURETHANE ON SPRAYED EPOXY LAYER
FRONT VIEW
HAZARDOUS AREA
APPROVAL LABEL
LOCATION
REAR TERMINAL
CONDUIT PORTS
FIBER OPTIC
BUSHING PORT
O-RING SEALED
RIGHT VIEW
3
LEFT VIEW
1
2
2
5
4
ENCLOSURE
JUNCTION BOX
1
3
4
5
6
7
EXTERNAL PROTECTIVE
CONDUCTOR TERMINAL
TERMINAL
COMPARTMENT
DOOR
7
6
MODEL NO. : XD-DR120WIN-4-N3-N3-N3-N3-M24/M5-SPU-CW
OR
XD-DR120WIN-4-M25-M25-M25-M25-M24/M5-SPU-CW
ATEX CERTIFICATION NO. : FTZÚ 08 ATEX 0182U
ATEX MARKING: II 2G Ex db IIC Gb
IEC Ex CERTIFICATION NO. : FTZÚ 09.0031U
IEC Ex MARKING: Ex db IIC Gb
Ԑx
JUNCTION BOX MANUFACTURER IS LIMATHERM S.A.:
CONTROL DRAWING: 2-Z-L2453
COATING: SPRAYED POLYURETHANE ON SPRAYED EPOXY LAYER
THE BOTTOM JUNCTION BOX PORT IS CONSUMED BY THE FIBER OPTIC SENSOR.
1/2" NPT FEMALE
FIBER OPTIC PORT/BUSHING THREADS ARE M24 X 1.5 6H/6g
ALL OXYvisor UNITS CONTAIN A BATTERY AND WILL HAVE LIVE CIRCUITS, EVEN
WHEN NO POWER IS APPLIED. DECLASSIFY HAZARDOUS AREAS BEFORE OPENING.
EXTERNAL PROTECTIVE
CONDUCTOR TERMINAL
Figure 15.2 - Class I Div. 2 HazLoc - Installation Control DWG - 2P0335 (p.2)
AC OPTION
DC OPTION
FRONT VIEW OF JUNCTION BOX
(DOOR REMOVED)
COMMUNICATION, TRANSMITTER, RELAY,
AND CONTROL BLOCK WIRING
1. ALL FIELD WIRING MUST HAVE 300V INSULATION, MINIMUM.
POWER WIRING: 12 TO 14 AWG, STRANDED COPPER CONDUCTOR.
COMMUNICATION, TRANSMITTER, RELAY AND CONTROL WIRING:
16 TO 24 AWG, STRANDED COPPER CONDUCTOR.
POWER
INPUT
WIRING
BLOCK
COMMUNICATION, TRANSMITTER, RELAY,
AND CONTROL BLOCK WIRING
POWER
INPUT
WIRING
BLOCK
2 OF 3
C. SHIELDS
OXYvisor HAZARDOUS LOCATION
REAR TERMINAL BOARD VIEWS
UNLESS
OTHERWISE
SPECIFIED
ALL ± .015"
FRACTIONAL
.000 ± .005"
.00 ± .010"
.0 ± .015"
APVD BY
DRAWN BY
DRAWING NO
DATE
SCALE
SIZE
SHEET
PART NO.
MATERIAL/DESCRIPTION
APVD
DWN
DESCRIPTION
DATE
REV
REVISIONS
ANGLES ± 1/4°
TOLERANCES
DIMENSIONS
IN INCHES
ECO#
5200 Convair Drive Carson City, NV 89706 PH(775)883-2500 FAX (775)297-4740
B. WOLF
01/26/2018
2P0335
-
OXYvisor
PART NO.
D
NTS
FOR
REAR VIEW OF ENCLOSURE
TERMINAL COMPARTMENT
(DOOR REMOVED)
INTERNAL PROTECTIVE
CONDUCTOR TERMINAL
INTERNAL PROTECTIVE
CONDUCTOR TERMINAL
FIBER OPTIC
CONNECTION
RTD
CONNECTION
A 09/26/18 U.L. CHANGES CS BW
EXTERNAL PROTECTIVE
CONDUCTOR TERMINAL