6-60 | Model 931S / Model 932S UV Analyzers
Decimal Alarm
Bit
Value
Type
Alarm Condition / Description and Suggested Corrective Action
4
16
Warning
w: Bench auto-setup problem
This condition indicates problems with the operation of the Optical Bench during
an Auto-Setup.
This alarm will be set if the Auto-Setup cannot hit
Signal Max
; the Auto-Setup will
be aborted.
This alarm will also be set if the
PMT Level
or
PMT Balance
value is greater
than 8000 mV or less than 2000 mV.
Corrective Action:
From the
Optical Bench
dialog box, check for deviations outside of these
ranges.
Perform the following checks, correct any faults, and restart the Auto-Setup:
•
Ensure that the Measuring Cell was zeroed.
Take appropriate safety precautions, open the Upper Enclosure, and:
•
Make sure the source lamps are not burned out.
•
Ensure the lamp sockets are attached securely to the lamps.
•
Ensure the lamps are inserted completely into the Lens Block in the Detector
Block Assembly.
•
Make sure the lamp power supply cables are connected and not damaged.
•
Ensure the PMT high-voltage (
JP100
,
JP101
) and flat (
J103
) cables are
connected and not damaged.
•
Align the source lamps and restart the Auto-Setup.
•
Adjust the jumper positions located at
P300
and
P301
on the
Optical Bench
board. See “Manipulating the PMT Level and PMT Balance” and Figure 6-10 in
this chapter for a description of the effects on the PMT Balance and PMT Level
in relation to different jumper positions.
•
Ensure the Measuring Cell and its Windows are clean and not damaged.
•
Check and clean, if necessary, the Mirror in the PMT Block.
Contact AMETEK
to verify operation before checking.
Setup (tab)
Optical
Bench
Содержание 931S
Страница 24: ...1 8 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Страница 34: ...2 10 Model 931S Model 932S UV Analyzers...
Страница 35: ...Specifications 2 11...
Страница 36: ...2 12 Model 931S Model 932S UV Analyzers...
Страница 37: ...Specifications 2 13...
Страница 38: ...2 14 Model 931S Model 932S UV Analyzers...
Страница 39: ...Specifications 2 15...
Страница 46: ...2 22 Model 931S Model 932S UV Analyzers...
Страница 71: ...Installation and Start Up 3 25 Figure 3 8 Conduit entries and AC power connection locations Lower Enclosure...
Страница 96: ...3 50 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Страница 170: ...4 74 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Страница 184: ...5 14 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Страница 221: ...Maintenance and Troubleshooting 6 37 Figure 6 12 Optical Bench Chopper Assembly Maintenance diagram 100 2313 1A...
Страница 237: ...Maintenance and Troubleshooting 6 53 Figure 6 14 AC Distribution board P N 100 1771...
Страница 238: ...6 54 Model 931S Model 932S UV Analyzers Figure 6 15 Host Controller board P N 100 1757 SW100...
Страница 239: ...Maintenance and Troubleshooting 6 55 Figure 6 16 Microcontroller board P N 100 1781 SW400...
Страница 250: ...6 66 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Страница 263: ...Appendix A Drawings A 7 RS 232 Cable Communications Connections 300 9480...
Страница 264: ...A 8 Model 931S Model 932S UV Analyzers Optical Bench Board 100 1662...
Страница 265: ...Appendix A Drawings A 9 PMT Buffer Board 100 0140...
Страница 268: ...S 2 Model 931S Model 932S UV Analyzers This page intentionally left blank...