6-46 | Model 931S / Model 932S UV Analyzers
Setup (tab)
Temp
Control
Decimal Alarm
Bit
Value
Type
Alarm Condition / Description and Suggested Corrective Action
2
4
Warning
w: Probe temperature range
The probe temperature has exceeded its
Too Hot
setpoint.
Note: This alarm is used only by analyzers using an unheated sample probe
(not controlling the sample probe).
Corrective Action:
•
From the
Temperature Control
dialog box, check the
Probe
Setpoint
and
Too Hot
setpoint. The
Probe
Setpoint
should be 5 °C to 10 °C above the
Too
Hot
temperature.
Take appropriate safety precautions, and:
•
Check the Probe RTD – inside the Probe Heater Assembly – for a short or
open circuit.
Contact AMETEK to verify operation before removing this
component.
3
8
Warning
w: Excessive zero drift
This condition indicates excessive Zero Drift, such as when Zero gas
transmission values attained after a Zero have exceeded user specified values.
Corrective Action:
•
Perform a Zero. See Chapter 5 for calibration information.
•
From the
Analyzer Data
tab, check the
Bench
Temperature
to ensure it has
stabilized.
•
From the
Status
tab, check the
HC Scode
and
MC Scode
alarm lists for
alarms that indicate contaminated Measuring Cell, optics, and/or sample
system.
If necessary, clean the appropriate component(s) of the sample system that
require cleaning.
•
From the
Gas Calibration
dialog box, ensure the
ZCalDriftSpec
values are
appropriate and have not been changed.
•
This condition may also indicate that the lamps need to be replaced, especially
if a “
w: ALC out of range
” alarm is also displayed on the
Status
tab (in the
MC Scode
alarm list). See “Source Lamp Replacement” in this chapter.
Setup (tab)
Gas
Calibration
Содержание 931S
Страница 24: ...1 8 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Страница 34: ...2 10 Model 931S Model 932S UV Analyzers...
Страница 35: ...Specifications 2 11...
Страница 36: ...2 12 Model 931S Model 932S UV Analyzers...
Страница 37: ...Specifications 2 13...
Страница 38: ...2 14 Model 931S Model 932S UV Analyzers...
Страница 39: ...Specifications 2 15...
Страница 46: ...2 22 Model 931S Model 932S UV Analyzers...
Страница 71: ...Installation and Start Up 3 25 Figure 3 8 Conduit entries and AC power connection locations Lower Enclosure...
Страница 96: ...3 50 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Страница 170: ...4 74 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Страница 184: ...5 14 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Страница 221: ...Maintenance and Troubleshooting 6 37 Figure 6 12 Optical Bench Chopper Assembly Maintenance diagram 100 2313 1A...
Страница 237: ...Maintenance and Troubleshooting 6 53 Figure 6 14 AC Distribution board P N 100 1771...
Страница 238: ...6 54 Model 931S Model 932S UV Analyzers Figure 6 15 Host Controller board P N 100 1757 SW100...
Страница 239: ...Maintenance and Troubleshooting 6 55 Figure 6 16 Microcontroller board P N 100 1781 SW400...
Страница 250: ...6 66 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Страница 263: ...Appendix A Drawings A 7 RS 232 Cable Communications Connections 300 9480...
Страница 264: ...A 8 Model 931S Model 932S UV Analyzers Optical Bench Board 100 1662...
Страница 265: ...Appendix A Drawings A 9 PMT Buffer Board 100 0140...
Страница 268: ...S 2 Model 931S Model 932S UV Analyzers This page intentionally left blank...