Calibration | 5-5
8. Bleed down the Span gas and disconnect it from the Span gas connec-
tion point.
9. Replace and tighten the Swagelok cap on the Span gas connection
point.
10. Turn on the Zero gas and adjust the pressure to the level it was at be-
fore Spanning the analyzer. Confirm that the pressure is at the recom-
mended pressure listed on the system drawings (in Documentation
Package shipped with the analyzer). The pressure must be less than
the 1380 kPag (200 psig) maximum Cell Pressure.
11. Change the
Flow Control
setting back to Analyzer Control by entering
‘
0
’ (
auto
) in the
Flow Control
text field
,
and then clicking
Apply
(next
to
Flow Control
text field). Click
OK
to return to the
Setup
tab, then
click
Save CFG to EEPROM
to permanently save the changes to the
analyzer’s EEPROM.
If the Flow Control mode has been in a mode other than
auto (‘0’) for
longer than the
AZero Interval (hr) duration, and you attempt to
return it to
auto (by entering ‘0’ and then clicking the Apply button
next to
Flow Control text field), the mode will not switch until after
the
Auto-Zero Timer (AZ-0) duration, under Auto Calibration
Timers, elapses.
If the Flow Control mode is left in any mode other than ‘
0’ (auto),
the message “
Flow Control not in Auto” will be displayed in the
Operational Message window on the Status tab.
NOTE
Figure 5-2.
Analyzer Data tab
(Model 932S).
Содержание 931S
Страница 24: ...1 8 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Страница 34: ...2 10 Model 931S Model 932S UV Analyzers...
Страница 35: ...Specifications 2 11...
Страница 36: ...2 12 Model 931S Model 932S UV Analyzers...
Страница 37: ...Specifications 2 13...
Страница 38: ...2 14 Model 931S Model 932S UV Analyzers...
Страница 39: ...Specifications 2 15...
Страница 46: ...2 22 Model 931S Model 932S UV Analyzers...
Страница 71: ...Installation and Start Up 3 25 Figure 3 8 Conduit entries and AC power connection locations Lower Enclosure...
Страница 96: ...3 50 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Страница 170: ...4 74 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Страница 184: ...5 14 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Страница 221: ...Maintenance and Troubleshooting 6 37 Figure 6 12 Optical Bench Chopper Assembly Maintenance diagram 100 2313 1A...
Страница 237: ...Maintenance and Troubleshooting 6 53 Figure 6 14 AC Distribution board P N 100 1771...
Страница 238: ...6 54 Model 931S Model 932S UV Analyzers Figure 6 15 Host Controller board P N 100 1757 SW100...
Страница 239: ...Maintenance and Troubleshooting 6 55 Figure 6 16 Microcontroller board P N 100 1781 SW400...
Страница 250: ...6 66 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Страница 263: ...Appendix A Drawings A 7 RS 232 Cable Communications Connections 300 9480...
Страница 264: ...A 8 Model 931S Model 932S UV Analyzers Optical Bench Board 100 1662...
Страница 265: ...Appendix A Drawings A 9 PMT Buffer Board 100 0140...
Страница 268: ...S 2 Model 931S Model 932S UV Analyzers This page intentionally left blank...