Maintenance and Troubleshooting | 6-61
Decimal Alarm
Bit
Value
Type
Alarm Condition / Description and Suggested Corrective Action
5
32
Warning
w: Sensor-Cal adjust problem
This condition indicates the sensor offset adjustment voltage has reached its
upper (90 %) or its lower (10 %) limit at the end of the sensor calibration.
This alarm will be set if, at the end of a sensor calibration, the sensor offset
voltage cannot be adjusted within the range of 10 % to 90 % in order to reach the
desired sensor
Zero-Cal Signal Setpoint
Target
value (sensor dependent).
Corrective Action:
•
From the
Sensors
dialog box, check the
Zero-Cal Signal Setpoint
Target
and
Delta
values for the specific sensor(s). Refer to the analyzer EEPROM
Data Sheets and check the factory-set
Signal ZCAL Target
(
Zero-Cal Signal
Setpoint
Target
) and
Signal ZCAL Delta
(
Zero-Cal Signal Setpoint
Delta
)
values for the specific sensor(s) installed. If necessary, adjust them to the
values listed on the EEPROM Data Sheets.
Take appropriate safety precautions, open the Electronics Enclosures and Oven,
and:
•
Check the cable between the sensor and
J109
on the
Customer I/O
board
proper connections and inspect it for damage (cuts, nicks, burn marks, etc.).
•
Measure the sensor signal voltage at the corresponding terminals for the
sensor, at
J109
on the
Customer I/O
board. For details on sensor wiring
assignments, refer to the Final “As-Built” Analyzer Customer Connections
drawing in the Analyzer Documentation Package.
If the sensor is faulty, replace the sensor.
Contact AMETEK to verify sensor
operation, to order a new sensor, and/or to remove and replace the
sensor.
•
Replace the
Micro-Interface
board.
Contact AMETEK to verify operation
before removing this board.
Setup (tab)
Sensors
Содержание 931S
Страница 24: ...1 8 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Страница 34: ...2 10 Model 931S Model 932S UV Analyzers...
Страница 35: ...Specifications 2 11...
Страница 36: ...2 12 Model 931S Model 932S UV Analyzers...
Страница 37: ...Specifications 2 13...
Страница 38: ...2 14 Model 931S Model 932S UV Analyzers...
Страница 39: ...Specifications 2 15...
Страница 46: ...2 22 Model 931S Model 932S UV Analyzers...
Страница 71: ...Installation and Start Up 3 25 Figure 3 8 Conduit entries and AC power connection locations Lower Enclosure...
Страница 96: ...3 50 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Страница 170: ...4 74 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Страница 184: ...5 14 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Страница 221: ...Maintenance and Troubleshooting 6 37 Figure 6 12 Optical Bench Chopper Assembly Maintenance diagram 100 2313 1A...
Страница 237: ...Maintenance and Troubleshooting 6 53 Figure 6 14 AC Distribution board P N 100 1771...
Страница 238: ...6 54 Model 931S Model 932S UV Analyzers Figure 6 15 Host Controller board P N 100 1757 SW100...
Страница 239: ...Maintenance and Troubleshooting 6 55 Figure 6 16 Microcontroller board P N 100 1781 SW400...
Страница 250: ...6 66 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Страница 263: ...Appendix A Drawings A 7 RS 232 Cable Communications Connections 300 9480...
Страница 264: ...A 8 Model 931S Model 932S UV Analyzers Optical Bench Board 100 1662...
Страница 265: ...Appendix A Drawings A 9 PMT Buffer Board 100 0140...
Страница 268: ...S 2 Model 931S Model 932S UV Analyzers This page intentionally left blank...