6-22 | Model 931S / Model 932S UV Analyzers
11. Click
on the
Standard Toolbar
to start scanning the
Measure
and
Reference
PMT signals.
Model 931S Analyzers:
• Observe the PMT signal of filter 1 if Lamp 1 was replaced.
• Observe the PMT signal of filter 2 if Lamp 2 was replaced.
Model 932S Analyzers:
(This filter assignment is the factory default for Model 932S Multi-Gas
Analyzers. See “Optical Bench” in Chapter 4 for other options.)
• Observe the PMT signals of filter positions 1, 3, and 5 if Lamp 1
was replaced.
• Observe the PMT signals of filter positions 2, 4, and 6 if Lamp 2
was replaced.
To change the Y-axis voltage scale, right-click over the graph. Click
Display and change the Upper limit and Lower limit.
Model 932S Analyzers only:
If the lamp pulse current-control value for a filter location is negative,
that filter location is not used.
12. Align the Source Lamps (see “Source Lamp Alignment” in this
chapter) to ensure optimal operation and then perform another
Auto-Setup
.
13. Swing the
Microcontroller/Micro-Interface
board assembly back to its
original position and tighten the (2) M3 X 8 screws on the mounting
bracket.
14. Change the
Flow Control
setting to ‘
2
’ (
zero
). Click
Apply
(next to
Flow Control
) to set the analyzer into
zero
Flow Control mode and
then click
OK
. Do not save the changes to EEPROM. Allow the Zero
gas to flow for 2–3 minutes.
15. From the
Status
tab, click
Conc. Alarm Soft Enable
On/Off
to
enable
the Concentration Alarms then click
OK
.
Change the
Flow Control
setting back to ‘
0
’ (
auto
). Click
OK
then
Apply
(next to
Flow Control
). Click
OK
then save this parameter
change to EEPROM by clicking the
Save CFG to EEPROM
button on
the
Setup
tab.
If the Flow Control mode is left in any mode other than ‘
0
’ (
auto
),
the message “
Flow Control not in Auto
” will be displayed in the
Operational Message
window on the
Status
tab.
NOTE
Содержание 931S
Страница 24: ...1 8 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Страница 34: ...2 10 Model 931S Model 932S UV Analyzers...
Страница 35: ...Specifications 2 11...
Страница 36: ...2 12 Model 931S Model 932S UV Analyzers...
Страница 37: ...Specifications 2 13...
Страница 38: ...2 14 Model 931S Model 932S UV Analyzers...
Страница 39: ...Specifications 2 15...
Страница 46: ...2 22 Model 931S Model 932S UV Analyzers...
Страница 71: ...Installation and Start Up 3 25 Figure 3 8 Conduit entries and AC power connection locations Lower Enclosure...
Страница 96: ...3 50 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Страница 170: ...4 74 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Страница 184: ...5 14 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Страница 221: ...Maintenance and Troubleshooting 6 37 Figure 6 12 Optical Bench Chopper Assembly Maintenance diagram 100 2313 1A...
Страница 237: ...Maintenance and Troubleshooting 6 53 Figure 6 14 AC Distribution board P N 100 1771...
Страница 238: ...6 54 Model 931S Model 932S UV Analyzers Figure 6 15 Host Controller board P N 100 1757 SW100...
Страница 239: ...Maintenance and Troubleshooting 6 55 Figure 6 16 Microcontroller board P N 100 1781 SW400...
Страница 250: ...6 66 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Страница 263: ...Appendix A Drawings A 7 RS 232 Cable Communications Connections 300 9480...
Страница 264: ...A 8 Model 931S Model 932S UV Analyzers Optical Bench Board 100 1662...
Страница 265: ...Appendix A Drawings A 9 PMT Buffer Board 100 0140...
Страница 268: ...S 2 Model 931S Model 932S UV Analyzers This page intentionally left blank...