4-68 | Model 931S / Model 932S UV Analyzers
Wheel Speed (rpm)
The current Optical Bench Chopper (Filter) Wheel speed in revolu-
tions per minute (RPM).
Model 932S Analyzers only.
Remote AZero
The current status of the remote, dry (potential free) contact, either
OFF
(closed) or
ON
(open). When
start
is displayed, the remote auto-
zero function has been initiated from a remote control room, such as a
Distributed Control System.
Warning Relay
The current status of the Warning Status
relay. When the analyzer is
operating in its
Normal
Run state, this relay is energized. This relay
is de-energized if a
Warning
alarm condition exists within the ana-
lyzer’s built-in diagnostic system (the analyzer requires attention). The
Warning contacts will also become de-energized if you manually force
the analyzer’s Flow Control Mode to any mode other than Analyzer
(
auto
) Control. When not in the
Normal state
, this field has a Red
background.
Resetting of the contacts is delayed by the
Sample Delay Timer
dura-
tion; otherwise, this relay’s contacts reset automatically upon correc-
tion of the
Warning
alarm or upon switching back to Analyzer Control
Mode.
Fault Relay
When the analyzer is operating in its
Normal
Run state, this relay is
energized. This relay is de-energized if a
Fault
alarm condition ex-
ists within the analyzer’s built-in diagnostic system (the analyzer
requires service). When not in the
Normal state
, this field has a Red
background.
When a
Fault
alarm condition exists (relay is de-energized), the sample
system Flow Control mode will automatically switch to “
zero
” mode
if the analyzer’s Flow Control Mode is set to “
auto
” (Analyzer Control
Mode).
If the relay is configured to latch (default) it will not clear unless you
manually reset the alarm; if not, it will reset automatically upon cor-
rection of the
Fault
alarm.
Whenever a
Fault
alarm condition exists, the text on the
Display
message lines on the
Optical Bench
,
Sensors
,
Gas Calibration
,
and
Information
dialog boxes will turn Red to indicate a fault status.
Содержание 931S
Страница 24: ...1 8 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Страница 34: ...2 10 Model 931S Model 932S UV Analyzers...
Страница 35: ...Specifications 2 11...
Страница 36: ...2 12 Model 931S Model 932S UV Analyzers...
Страница 37: ...Specifications 2 13...
Страница 38: ...2 14 Model 931S Model 932S UV Analyzers...
Страница 39: ...Specifications 2 15...
Страница 46: ...2 22 Model 931S Model 932S UV Analyzers...
Страница 71: ...Installation and Start Up 3 25 Figure 3 8 Conduit entries and AC power connection locations Lower Enclosure...
Страница 96: ...3 50 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Страница 170: ...4 74 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Страница 184: ...5 14 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Страница 221: ...Maintenance and Troubleshooting 6 37 Figure 6 12 Optical Bench Chopper Assembly Maintenance diagram 100 2313 1A...
Страница 237: ...Maintenance and Troubleshooting 6 53 Figure 6 14 AC Distribution board P N 100 1771...
Страница 238: ...6 54 Model 931S Model 932S UV Analyzers Figure 6 15 Host Controller board P N 100 1757 SW100...
Страница 239: ...Maintenance and Troubleshooting 6 55 Figure 6 16 Microcontroller board P N 100 1781 SW400...
Страница 250: ...6 66 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Страница 263: ...Appendix A Drawings A 7 RS 232 Cable Communications Connections 300 9480...
Страница 264: ...A 8 Model 931S Model 932S UV Analyzers Optical Bench Board 100 1662...
Страница 265: ...Appendix A Drawings A 9 PMT Buffer Board 100 0140...
Страница 268: ...S 2 Model 931S Model 932S UV Analyzers This page intentionally left blank...