3-9
Parameter Setup
6. View 3050-AP Analyzer parameter readings using the Monitor Tab. Refer to figure 3.9.
From this tab you can also collect data, test the alarm contacts and mA output of the
analyzer.
Using the test buttons takes the analyzer offline.
The Test Alarms push buttons allow you to toggle the alarm contacts to an opened or
closed state. Use a multi-meter set to ohms to read resistance. Refer to figure 2.5 for
location of contacts on analyzer.
• Opened contacts should read infinity.
• Closed contacts should read zero.
The Test mA Output push buttons allow you to test the analog outputs. Use a meter
to test output. Refer to figure 2.6 of for wiring.
To discontinue test mode, press another tab. Test mode will automatically time out after
10 minutes of inactivity.
When analyzer switches from test mode to online, the analyzer resets itself.
NOTE
Figure 3.9. AMETEK 3050-AP Configurator
- Monitor Tab
(The AP Analyzer does not support the Flow Adjust option.)
NOTE
Содержание 3050-AP
Страница 10: ...x 3050 AP Moisture Analyzer This page intentionally left blank...
Страница 19: ...Instalation and Startup 2 3 Figure 2 1 Model 3050 AP front and side view...
Страница 20: ...2 4 3050 AP Moisture Analyzer Figure 2 2 Model 3050 AP rear view...
Страница 24: ...2 8 3050 AP Moisture Analyzer Figure 2 5 3050 AP Power and Signal Connections...
Страница 25: ...Instalation and Startup 2 9 Figure 2 6 4 20 mA Output Wiring...
Страница 26: ...2 10 3050 AP Moisture Analyzer Figure 2 7 RS 232 Wiring...
Страница 27: ...Instalation and Startup 2 11 Figure 2 8 RS 485 Cables Multiple 3050 OLV s...
Страница 28: ...2 12 3050 AP Moisture Analyzer Figure 2 9 RS 485 to RS 232 Conversion for host PC...
Страница 46: ...5 4 3050 AP Moisture Analyzer This page intentionally left blank...