3-6
3050-AP Moisture Analyzer
Sensor Saver
Not Checked
Analyzer operates with a rapid cycle time, minimizing system
response time.
Checked
Analyzer operates with a slow cycle time, maximizing cell life at
the expense of system response time.
4 to 20 mA output Outputs moisture concentration .
20 mA
Enter upper range output value.
4 mA
Enter lower range output value.
Hold during verify Check to hold moisture output at last measured value during
and zero
verification and zero.
Alarm Output
Enable
Check to activate alarm.
High limit
Enter high limit value
Low limit
Enter low limit value
4. Define Dryer and Clock Settings using the Verify Tab. Refer to figure 3.5.
• Dryer Production Code Enter dryer production code number when installing
dryer.
• Set Clock
Press button to set clock on analyzer.
• Adjust Span
Enter new span value. Use the following sequence to
obtain new value.
1. Change span to 1.0
2. Inject gas of a known moisture concentration into sample inlet until reading
stablizes. View concentration reading on Status Tab.
3. Calulate new span value.
Actual gas concentration
3050 concentration reading (Status Tab)
Example:
50 ppmv
45 ppmv
4. Enter new span value and press enter.
= new span
= 1.11
Содержание 3050-AP
Страница 10: ...x 3050 AP Moisture Analyzer This page intentionally left blank...
Страница 19: ...Instalation and Startup 2 3 Figure 2 1 Model 3050 AP front and side view...
Страница 20: ...2 4 3050 AP Moisture Analyzer Figure 2 2 Model 3050 AP rear view...
Страница 24: ...2 8 3050 AP Moisture Analyzer Figure 2 5 3050 AP Power and Signal Connections...
Страница 25: ...Instalation and Startup 2 9 Figure 2 6 4 20 mA Output Wiring...
Страница 26: ...2 10 3050 AP Moisture Analyzer Figure 2 7 RS 232 Wiring...
Страница 27: ...Instalation and Startup 2 11 Figure 2 8 RS 485 Cables Multiple 3050 OLV s...
Страница 28: ...2 12 3050 AP Moisture Analyzer Figure 2 9 RS 485 to RS 232 Conversion for host PC...
Страница 46: ...5 4 3050 AP Moisture Analyzer This page intentionally left blank...