PN 903-8797, Rev A
The User Interface is made up of two areas:
•
Display Screen/LEDs
Consists of one self-illuminated line capable of displaying up to 12 alpha-
numeric characters and one row of seven LEDs. The information displayed
depends on the current operation mode.
- The self-illuminated line displays the results of the most recent measure-
ment cycle, mirror temperature, run time, and Status Code (SCode).
To change an operating parameter, it must be active on the display. The
display line flashes when the 241CE II Analyzer is in Standby Mode (a
Fault alarm is present).
- The LEDs provide a visual indicator of the built-in diagnostics function.
When a LED is On, it indicates the alarm condition associated with the
LED currently exists. For more information about this diagnostics tool,
see “Types of Alarms” in Chapter 4.
The LEDs, numbered 1–7 from left to right, and their conditions when
turned On are:
LED #
Setpoint By
Condition Existing When LED Turned On
1
User
The Hydrocarbon Dewpoint temperature Off-Spec
Alarm is disabled (
Stat\History : En\Dis Alrm?
).
2
Diagnostics
Fault Alarm: Analyzer in Standby Mode –
Immediate
Service Required
(review
SCode
and
XSCode
in the
Stat\History
menu to gather information before
contacting AMETEK). The alphanumeric display will
also flash to alert the operator to this condition.
3
Diagnostics
Warning Alarm: Analyzer reporting an abnormal
condition –
Investigation Required
(review
SCode
and
XSCode
in the
Stat\History
menu and attempt to
correct the problems).
4
Diagnostics
Hydrocarbon Dewpoint Temperature Not Found (
NoD
),
or Purging/Hold Stage Is In Progress (
Pur
). This setting
is made in the
ApplicCFG : UsrRelay
menu.
See also the
UsrRelay
description under “Working in
the Application Configuration Menu” in this chapter.
5
User
Reserved for Water Content OffSpec Alarm setpoint.
Currently Not Used.
6
User
The current Hydrocarbon Dewpoint temperature is
higher than the Hydrocarbon Dewpoint temperature
Off-Spec Alarm setpoint value (
ApplicCFG : HCDPA
).
This alarm will only be seen if the Off-Spec Alarm is
enabled in the Stat\History menu (LED 1 is Off when
the Off-Spec Alarm is enabled).
7
User
Reserved for Water Dewpoint temperature
OffSpec
Alarm
setpoint.
Currently Not Used.
3-2 | 241CE II Hydrocarbon Dewpoint Analyzer
Содержание 241CE II
Страница 14: ...PN 903 8797 Rev A This page intentionally left blank xiv 241CE II Hydrocarbon Dewpoint Analyzer...
Страница 19: ...PN 903 8797 Rev A Figure 1 3 Measuring Cell Assembly cutaway view Overview 1 5...
Страница 24: ...PN 903 8797 Rev A This page intentionally left blank 1 10 241CE II Hydrocarbon Dewpoint Analyzer...
Страница 32: ...PN 903 8797 Rev A Figure 2 1 Analyzer backpan mounting details 2 8 241CE II Hydrocarbon Dewpoint Analyzer...
Страница 66: ...PN 903 8797 Rev A This page intentionally left blank 2 42 241CE II Hydrocarbon Dewpoint Analyzer...
Страница 120: ...PN 903 8797 Rev A Figure 4 3 Window Push button assembly details 4 32 241CE II Hydrocarbon Dewpoint Analyzer...
Страница 143: ...PN 903 8797 Rev A Figure 4 6 Measuring Cell Assembly components Maintenance and Troubleshooting 4 55...
Страница 150: ...PN 903 8797 Rev A This page intentionally left blank 5 6 241CE II Hydrocarbon Dewpoint Analyzer...
Страница 164: ...PN 903 8797 Rev A Analyzer Block Diagram WX 241CE II 1 B 2 241CE II Hydrocarbon Dewpoint Analyzer...
Страница 165: ...PN 903 8797 Rev A Digital Communications Wiring Diagram WX 241CE II 5 Appendix B B 3...
Страница 170: ...PN 903 8797 Rev A This page intentionally left blank B 8 241CE II Hydrocarbon Dewpoint Analyzer...