Fault Alarms
Fault alarms indicate that a serious condition has been detected which inter-
feres with proper operation of the analyzer – a failure of an integral part of the
system has occurred. When a Fault alarm is present, the alphanumeric display
will flash and LED 2 turns On to alert the operator that immediate attention is
required (the analyzer requires service).
Because any results obtained under a Fault alarm condition are invalid, the
operation of the analyzer is suspended and the analyzer is placed into Standby
Mode.
Off-Spec Alarms
An Off-Spec – or out-of-specification – Alarm indicates that the current value
for the
Off-Spec
parameter is higher than a preset value. The Off-Spec Alarm
can be enabled from the
Stat\History
menu.
The Off-Spec Alarm currently is used only for HCDP. The HCDP Off-Spec Alarm
(
HCDPA
) indicates that the current HCDP temperature is above (greater than)
the HCDP temperature Off-Spec Alarm setpoint value (
ApplicCFG : HCDPA
).
When the Hydrocarbon Dewpoint Off-Spec Alarm (
HCDPA
) is triggered, LED 6
on the User Interface turns On, but only if the Off-Spec Alarm is enabled (
Stat\
History : En\Dis Alrm?
). If triggered, this alarm will be also be displayed on the
User Interface if the Off-Spec Alarm is enabled.
4-38 | 241CE II Hydrocarbon Dewpoint Analyzer
Содержание 241CE II
Страница 14: ...PN 903 8797 Rev A This page intentionally left blank xiv 241CE II Hydrocarbon Dewpoint Analyzer...
Страница 19: ...PN 903 8797 Rev A Figure 1 3 Measuring Cell Assembly cutaway view Overview 1 5...
Страница 24: ...PN 903 8797 Rev A This page intentionally left blank 1 10 241CE II Hydrocarbon Dewpoint Analyzer...
Страница 32: ...PN 903 8797 Rev A Figure 2 1 Analyzer backpan mounting details 2 8 241CE II Hydrocarbon Dewpoint Analyzer...
Страница 66: ...PN 903 8797 Rev A This page intentionally left blank 2 42 241CE II Hydrocarbon Dewpoint Analyzer...
Страница 120: ...PN 903 8797 Rev A Figure 4 3 Window Push button assembly details 4 32 241CE II Hydrocarbon Dewpoint Analyzer...
Страница 143: ...PN 903 8797 Rev A Figure 4 6 Measuring Cell Assembly components Maintenance and Troubleshooting 4 55...
Страница 150: ...PN 903 8797 Rev A This page intentionally left blank 5 6 241CE II Hydrocarbon Dewpoint Analyzer...
Страница 164: ...PN 903 8797 Rev A Analyzer Block Diagram WX 241CE II 1 B 2 241CE II Hydrocarbon Dewpoint Analyzer...
Страница 165: ...PN 903 8797 Rev A Digital Communications Wiring Diagram WX 241CE II 5 Appendix B B 3...
Страница 170: ...PN 903 8797 Rev A This page intentionally left blank B 8 241CE II Hydrocarbon Dewpoint Analyzer...