PN 903-8797, Rev A
Filterblock Maintenance
The Membrane Filters in the Filterblock Assembly should be inspected periodi-
cally for signs of contamination or damage.
Clean the Filterblock:
• If alarms that indicate the optical paths show signs of contamination are
displayed, such as:
-
f H2O Optics
(SCode 32)
-
f HC Optics
(SCode 64)
• If there is evidence of baseline deviations (see “Checking and Recording
the Baseline Settings” in the Diagnostic Flowcharts section of this chapter).
• Each time you clean the Measuring Cell.
• If the internal parts are severely contaminated, more frequent cleaning
may be required.
Replace parts in the Filterblock:
• If, while cleaning the Filterblock, the Membrane Filters are found to be
damaged (are creased, have holes, or are transparent) or severely contami-
nated (with liquids, oils, or waxes).
See “Spare Part Lists”
in this chapter for ordering
numbers.
RUN Mode
Stat\History
SCode
4-22 | 241CE II Hydrocarbon Dewpoint Analyzer
Содержание 241CE II
Страница 14: ...PN 903 8797 Rev A This page intentionally left blank xiv 241CE II Hydrocarbon Dewpoint Analyzer...
Страница 19: ...PN 903 8797 Rev A Figure 1 3 Measuring Cell Assembly cutaway view Overview 1 5...
Страница 24: ...PN 903 8797 Rev A This page intentionally left blank 1 10 241CE II Hydrocarbon Dewpoint Analyzer...
Страница 32: ...PN 903 8797 Rev A Figure 2 1 Analyzer backpan mounting details 2 8 241CE II Hydrocarbon Dewpoint Analyzer...
Страница 66: ...PN 903 8797 Rev A This page intentionally left blank 2 42 241CE II Hydrocarbon Dewpoint Analyzer...
Страница 120: ...PN 903 8797 Rev A Figure 4 3 Window Push button assembly details 4 32 241CE II Hydrocarbon Dewpoint Analyzer...
Страница 143: ...PN 903 8797 Rev A Figure 4 6 Measuring Cell Assembly components Maintenance and Troubleshooting 4 55...
Страница 150: ...PN 903 8797 Rev A This page intentionally left blank 5 6 241CE II Hydrocarbon Dewpoint Analyzer...
Страница 164: ...PN 903 8797 Rev A Analyzer Block Diagram WX 241CE II 1 B 2 241CE II Hydrocarbon Dewpoint Analyzer...
Страница 165: ...PN 903 8797 Rev A Digital Communications Wiring Diagram WX 241CE II 5 Appendix B B 3...
Страница 170: ...PN 903 8797 Rev A This page intentionally left blank B 8 241CE II Hydrocarbon Dewpoint Analyzer...