Short instructions
The following short instructions are intended as a guide to users
who are fully familiar with safety precautions and operating
instructions described in this manual. The instructions assumes
that the instrument is installed according to the installation
instructions.
Calibrating
Please wait...
1
Switch on the instrument
by means of the mains
switch on the rear panel.
2 The
main operating menu
(RUN-menu) is shown.
3
Calibrate
the pH electrode before use and/or daily
by using 2 buffers with known pH values.
4 The
conductivity flow cell
does not normally need
to be calibrated.
5 Set the
conductivity scaling
for 0 and 100%.
6 Always store the
pH electrode
in 1:1 mixture of
pH 4 buffer and 1 M KNO
3
when not in use.
Select
OK
ESC
pH/C-900
Calibrate pH
(7.00 – 10.00)
Cond Temp pH
25.4% 22.9C 11.5
Set Cond Scale 100 %
(50.00 mS/cm)
TC information ab, Uppsala. Printed in Sweden by TK i Uppsala AB
18-1120-06_AC.qxd 1999-04-14 14:16 Sida 36 (1,1)
Содержание AKTA design pH/C-900
Страница 1: ...User Manual 18 1120 06 Monitor pH C 900 18 1120 06_AC qxd 1999 04 14 14 16 Sida 1 1 1...
Страница 36: ...18 1120 06_AC qxd 1999 04 14 14 16 Sida 32 1 1...
Страница 37: ...18 1120 06_AC qxd 1999 04 14 14 16 Sida 33 1 1...
Страница 38: ...April 1999 18 1120 06_AC qxd 1999 04 14 14 16 Sida 34 1 1...
Страница 39: ...18 1120 06_AC qxd 1999 04 14 14 16 Sida 35 1 1...