A.2 pH electrode
The pH electrode is of the sealed
combination double junction type.
It contains a sealed Ag/AgCl
reference which cannot be refilled,
an internal electrolyte bridge of 4M
KCl saturated with Ag/AgCl, an
outer electrolyte bridge of 1M
KNO
3
, an annular ceramic
reference junction and a low profile
pH membrane. The pH electrode is
delivered with a transparent cover.
The flow cell should not be used
with any other pH electrode.
A.3 Conductivity cell
The flow cell has two cylindrical titanium electrodes positioned in
the flow path of the cell. An alternating voltage is applied between
the electrodes and the resulting current is measured and used to
calculate the conductivity of the eluent. The monitor controls the AC
frequency and increases it with increasing conductivity between 50
Hz and 50 kHz giving maximum linearity and true conductivity
values.
The conductivity is automatically calculated by multiplying the
measured conductance by the flow cell’s cell constant. The cell
constant is pre-calibrated on delivery but can be measured with a
separate calibration procedure. This procedure is described in
Reference information section B.2.
One of the electrodes has a small temperature sensor for measuring
the temperature of the eluent in the flow cell. Temperature variations
influence the conductivity and in some applications, when highly
precise conductivity values are required, it is possible to program a
temperature compensation factor that recalculates the conductivity
to a set reference temperature.
Monitor pH/C-900
User Manual 18-1120-06 Edition AC
20
Eluent
Temperature sensor
Sealing
Internal electrolyte
bridge containing
4M KCl saturated
with AgCl
Silicone sealing
Internal annular
coaxial ceramic
reference junction
External electrolyte
bridge containing
1M KNO
3
O-ring
Glass electrode
containing diluted,
buffered KCl
External annular
coaxial ceramic
reference junction
O-ring
Ag/AgCl reference
electrode
A
Reference information
18-1120-06_AC.qxd 1999-04-14 14:16 Sida 20 (1,1)
Содержание AKTA design pH/C-900
Страница 1: ...User Manual 18 1120 06 Monitor pH C 900 18 1120 06_AC qxd 1999 04 14 14 16 Sida 1 1 1...
Страница 36: ...18 1120 06_AC qxd 1999 04 14 14 16 Sida 32 1 1...
Страница 37: ...18 1120 06_AC qxd 1999 04 14 14 16 Sida 33 1 1...
Страница 38: ...April 1999 18 1120 06_AC qxd 1999 04 14 14 16 Sida 34 1 1...
Страница 39: ...18 1120 06_AC qxd 1999 04 14 14 16 Sida 35 1 1...