B.2.4 Setting up pH temperature compensation
The relationship between pH and the output signal from the pH
electrode is temperature dependent. For more accurate measurements
during temperature changes, the pH measurement can be
temperature compensated. In normal applications, if the temperature
of the buffers and calibration buffers are identical, temperature
compensation does not need to be on.
When using temperature compensation it is important that the
temperature of the pH electrode is the same as that of the
conductivity flow cell since that is where the temperature is
measured.
1 Select main menu
Setup
, press OK.
2 Select menu
Setup pH
, press OK.
3 Select sub menu
Setup pH Temp Comp
, press OK.
4 Set the temperature compensation on or off, press OK.
B.2.5 Setting up the pH display
Normally the pH is displayed in the main operating menu or its
alternative (see section 3.4 Reading pH and conductivity values). If
not required the pH display can be set to off.
1 Select main menu
Setup
, press OK.
2 Select menu
Setup pH
, press OK.
3 Select sub menu
Setup Show pH
, press OK.
4 Set the pH display on or off, press OK.
B.2.6 Calibrating the temperature sensor
Calibration of the temperature sensor in the conductivity flow cell is
only necessary if the monitor is used in high accuracy measurement
or the conductivity flow cell is replaced.
1 Place the flow cell together with a precision thermometer inside a
box or empty beaker to ensure that they are not exposed to draft.
Leave them for 15 minutes to let the temperature stabilise.
2 Read the temperature on the thermometer.
3 Select main menu
Setup
, press OK.
4 Select sub menu
Setup Temperature
, press OK.
5 Select sub menu
Setup Adjust Temp
, press OK.
6 A warning message will be displayed, press OK.
7 Enter the temperature shown on the thermometer, press OK.
Monitor pH/C-900
User Manual 18-1120-06 Edition AC
24
B
Reference information
Warning! This will
change temp calib
Setup Adjust Temp
(23.5°C)
Setup Temperature
Setup Adjust Temp
(23.5°C) 25.0
Setup Show pH
(On)
Setup pH Temp Comp
(Off) On Off
Setup pH
18-1120-06_AC.qxd 1999-04-14 14:16 Sida 24 (1,1)
Содержание AKTA design pH/C-900
Страница 1: ...User Manual 18 1120 06 Monitor pH C 900 18 1120 06_AC qxd 1999 04 14 14 16 Sida 1 1 1...
Страница 36: ...18 1120 06_AC qxd 1999 04 14 14 16 Sida 32 1 1...
Страница 37: ...18 1120 06_AC qxd 1999 04 14 14 16 Sida 33 1 1...
Страница 38: ...April 1999 18 1120 06_AC qxd 1999 04 14 14 16 Sida 34 1 1...
Страница 39: ...18 1120 06_AC qxd 1999 04 14 14 16 Sida 35 1 1...