Calibrating pH (optional in ÄKTApurifier systems)
A good laboratory routine is to calibrate the instrument once a day,
when the electrode is replaced and if the ambient temperature is
changed. The pH monitor is calibrated using standard buffer
solutions in a two point calibration. The two buffer solutions can
have any pH value as long as the difference between them is at least
1 pH unit. Calibration can also be performed from UNICORN. In
UNICORN select
System Control:System:Calibrate
. Select the pH
monitor. The calibration procedure can be done with the pH
electrode either fitted in or removed from the flow cell.
Calibrating with the electrode outside the flow cell
When calibrating the electrode out of the flow cell and changing
from one buffer to another, rinse the electrode tip with distilled
water and dab it carefully with a soft tissue to absorb the remaining
water. Do NOT wipe the electrode as this may charge it and give
unstable readings.
The steps below describe the procedure used with the electrode
removed from the flow cell.
Note: The Monitor must be unlocked if connected to a UNICORN
control system.
1 Remove the pH electrode from the flow cell and immerse the
electrode in the 1:st standard buffer solution (normally pH 7.0).
2 Select main menu
Calibrate pH
. The display shows the current
low and high calibrated pH value. Press OK.
3 Select sub menu
Calib pH Buffer 1
,
press OK. When the pH value
has stabilized, the
Please wait
message will disappear.
4 Adjust the pH value in the display using the dial, so that it
corresponds to the known pH value of the 1:st buffer solution,
press OK. The sub menu
Calib pH Buffer 2
is shown.
5 Rinse the electrode tip with distilled water and then immerse the
electrode in the 2:nd standard buffer solution (e.g. 4.0 or 9.0),
press OK.
6 When the pH value has stabilized, the
Please wait
message will
disappear.
7 Adjust the pH value in the display using the dial, so that it
corresponds to the known pH value of the 2:nd buffer solution,
press OK.
Monitor pH/C-900
User Manual 18-1120-06 Edition AC
9
Operation
3
Calib pH Buffer 1
Calib pH Buffer 2
Calibrate pH
(7.00 – 10.00)
18-1120-06_AC.qxd 1999-04-14 14:16 Sida 9 (1,1)
Содержание AKTA design pH/C-900
Страница 1: ...User Manual 18 1120 06 Monitor pH C 900 18 1120 06_AC qxd 1999 04 14 14 16 Sida 1 1 1...
Страница 36: ...18 1120 06_AC qxd 1999 04 14 14 16 Sida 32 1 1...
Страница 37: ...18 1120 06_AC qxd 1999 04 14 14 16 Sida 33 1 1...
Страница 38: ...April 1999 18 1120 06_AC qxd 1999 04 14 14 16 Sida 34 1 1...
Страница 39: ...18 1120 06_AC qxd 1999 04 14 14 16 Sida 35 1 1...