Faults and actions
If the suggested actions do not correct the fault, call Amersham
Pharmacia Biotech.
pH measurement (optional in ÄKTApurifier systems)
Fault
Action
No response to pH changes
1
Check that the electrode cable is connected properly to
the rear of the instrument.
2
The electrode glass membrane may be cracked. Replace
the electrode.
Small response to pH changes
1
Clean the pH electrode according to section
4.3 and
recalibrate.
2
If the problem remains, replace the pH electrode.
1
Check the pH electrode tip. If it is contaminated
clean the electrode following the instructions in section
4.3 Cleaning the pH electrode.
2
If the membrane has dried out, the electrode may be
restored by soaking it in buffer overnight.
3
Clogged liquid junction. Refer to section
4.3 Cleaning the pH
electrode.
Incorrect/ unstable pH reading
1
Check that the electrode cable is connected properly to
the rear of the instrument.
2
Check that the pump and valves operates correctly.
3
Check that the electrode is correctly inserted in the flow cell
and, if necessary, hand-tighten the nut.
4
If air in the flow cell is suspected, tap the flow cell carefully
or tilt it to remove the air. Alternatively flush the flow cell with
buffer at 8 ml/min for 1/2 min.
5
Check that the pH electrode is not broken.
6
Check that the pH electrode is calibrated.
Monitor pH/C-900
User Manual 18-1120-06 Edition AC
15
Trouble shooting
5
Monitor pH/C900
V1.00
Slow pH response
or
Calibration impossible
WARNING! The instrument must not be opened by the user. It
contains high voltage circuits which can be capable of delivering a
lethal electric shock.
18-1120-06_AC.qxd 1999-04-14 14:16 Sida 15 (1,1)
Содержание AKTA design pH/C-900
Страница 1: ...User Manual 18 1120 06 Monitor pH C 900 18 1120 06_AC qxd 1999 04 14 14 16 Sida 1 1 1...
Страница 36: ...18 1120 06_AC qxd 1999 04 14 14 16 Sida 32 1 1...
Страница 37: ...18 1120 06_AC qxd 1999 04 14 14 16 Sida 33 1 1...
Страница 38: ...April 1999 18 1120 06_AC qxd 1999 04 14 14 16 Sida 34 1 1...
Страница 39: ...18 1120 06_AC qxd 1999 04 14 14 16 Sida 35 1 1...