A P P L I C A T I O N A N D O P E R A T I O N
E N
Fig. 22
1
Gas flow directly after starting
2
Gas flow during the cooling phase
The correct quantity of protective gas must be
ensured prior to the first program run. It is manu-
ally set and checked with the
F1 and F2 functions.
▷
Open the valve of the gas cylinder.
▷
Access the configuration level with the
key.
▷
Press the
key.
▷
Press the
and/or
key.
▷
Press the
key.
The function starts.
▷
Upon completion of the test, press the
key to
end the function.
The manometer reading of the pressure reducer
for the gas cylinder must be 1 bar.
In case the value deviates:
▷
Make sure that the right-hand valve on the pres-
sure reducer is completely opened.
▷
Readjust to exactly 1 bar by turning the black
regulating knob of the pressure reducer.
Fig. 23
1
Valve
2
Regulating knob
After the adjustment, the furnace manometer
reading at
“
F1
”
must indicate a flow value of 0.2
bar and at
“
F2
”
a flow value of 0.9 bar.
1
2
1 bar
1
2