The machine is intended exclusively for use
within dry, closed rooms.
_
The machine must be positioned below an
exhaust pipe or an exhaust hood (as used for
pre-heating furnaces).
_
Sufficient ventilation must be ensured at the
set-up location, in order to draw off the heat and
possibly developing exhaust gas.
_
A clearance of at least 50 mm is to be observed
sidewards, to the rear and upwards of the fur-
nace. The lateral clearance to non-flammable
materials can be reduced to 20 cm. The space
requirements for the CoCr sintering furnace
(incl. connections) are thus:
▪
Width: 1.4 m (0.8 m for non-flammable materi-
als)
▪
Depth: 0.9 m
▪
Height: 1.0 m
_
The set-up surface must be level.
_
The empty weight of the machine is 30 kg. The
set-up surface must withstand an appropriate
load.
_
During operation, the room temperature must
be between 5
°C
and 40
°C
; avoid large varia-
tions in temperature.
_
The floor covering must be made of non-flam-
mable material, so that hot material falling out
of the furnace can not inflame the covering.
▷
Set up the machine on a sturdy work table or
work bench (on a non-flammable surface).
▷
Remove all packaging material, including from
the furnace compartment.
NOTE:
Malfunction!
▷
Do not install an exhaust pipe directly near the
exhaust air pipe of the furnace!