Specifications*
Size
3.88
” (9.86cm) Wide
7.72
” (19.61cm) Long
1.58
” (4.01cm) Deep in Hand-held region, 2.14”
(5.44cm) deep at the top end
Weight
1.1 pound (0.5 kg)
Operating Range
0 - 50 degrees C (32 - 122 degrees F)
Relative humidity to 70% (non-condensing)
Meets mechanical shock requirements of MIL-PRF
28800F for a class 1 (harshest) environment
Safety and Environmental Protection
CE rated 600V input CAT IV per IEC61010-1:2001
IP40 per IEC 529
Power Requirement
12 VDC @ 500 ma, wall mount power supply included
Internal Li-ion operates 8 hours after 4 hours charge.
Voltage
Input Range: 1 - 600 Vrms steady-state (direct input),
or 600
– 5,000 Vrms with 5KVP probes,
or 600
– 15,000 Vrms with 15KVP probes.
Display Range: 1 - 6 megavolts (using input ratios)
Accuracy:
0.1% of reading (over input range) +/- 0.3 Vrms
Measure Rate: Measure every cycle of every input.
Frequency Response: No de-rating of accuracy for
harmonics through 3900 Hz
Current
Input Range: 0.01 - 5000 Amps, AC or DC with the
proper current probe attached
Display Range: 1ma - 6 megamps (using input ratios)
Accuracy:
0.1% of reading plus accuracy of probe
Measure Rate: Measure every cycle of every input.
Frequency Response: dependent on current probe attached. With AT6001: no de-rating of
accuracy for harmonics through 3900 Hz)
Frequency
Range:
DC, 22 - 200 Hz, 360
– 440 Hz fundamental frequency
DC and 22 - 4500 Hz included in RMS
22 - 3900 Hz for harmonic measurements
Accuracy:
0.25%
Power, Energy, Cost, Power Factor
Display Range: 1 watt - 60 megawatts (using input ratios)
Accuracy:
0.5% plus accuracy of current probe
Measure Rate: Measure every cycle of every input.
Harmonic Distortion:
Range:
Display of THD and individual harmonics through 25th (1800 Hz) of all signals
Содержание ATPOL II